Selective reduction of multifetal pregnancy using high-intensity focused ultrasound in the rabbit model.
OBJECTIVE: High-order multifetal pregnancies carry a significant risk of obstetric complications and poor pregnancy outcome. Selective reduction has traditionally been performed using transabdominal and transvaginal ultrasound-guided intracardiac injection of potassium chloride. We have previously shown that high-intensity focused ultrasound (HIFU) can create a coagulative tissue necrosis in the sheep fetus. The objective of this study was to investigate the feasibility of non-invasive selective fetal reduction using HIFU in a rabbit model. METHODS: A protocol for HIFU-induced tissue coagulation was developed in the rabbit model. The fetal heart was targeted with ultrasound-guided tissue ablation by a HIFU beam. Five time-mated does between 20-29 days' gestation underwent transabdominal fetal cardiac ablation in a total of 11 fetuses. The HIFU system consisted of a 7-MHz high-power transducer, operated at 2000 W/cm2. The fetal heart rate was observed using real-time ultrasound with Doppler flow velocimetry. All lesions were assessed macroscopically and by histological analysis. RESULTS: Severe bradycardia leading to asystole was observed in all targeted fetuses with ultrasound examination. Dissection of fetuses demonstrated a necrotic intrathoracic lesion similar in size to the HIFU focus (approximately 1 x 9 mm). None of the surrounding fetuses was found to have bradycardia during the procedure or a macroscopic lesion on dissection. CONCLUSION: In this pilot study HIFU seems promising to ablate even highly vascularized tissue in the fetus.