PubMed Health⌕ Search

Biomedical subjects

Setsuo Takatani

Publications and source records attributed to Setsuo Takatani.

21 records · Page 2Linked to original sources

New method for the detection of thrombus formation in cardiovascular devices: optical sensor evaluation in a flow chamber model.

We proposed a new method of detecting the onset of thrombus formation based upon the backscattered light intensity changes caused by the alteration of blood flow behavior in the cardiopulmonary devices. In an optical senor based upon the First Order Scattering theory, the relationship between the backscattered light intensity and hematocrit exhibited a monotonically decaying curve for the hematocrit level greater than 45%. To distinguish the effect either by thrombus formation or erythrocyte aggregation on the back-scattered light intensity with a flow chamber connected to a rabbit arterial-venous bypass model, we generated an oscillatory flow using a roller pump and analyzed the results using a Fast Fourier Transform (FFT) method. Our hypothesis was that the nonclotting blood flow would yield an unchanged fundamental spectral power density of the oscillation frequency generated by a roller pump, whereas the thrombus formation would attenuate its power. We measured the back-scattered light intensity in the flow chamber of high shear region and low shear region (n = 5). The blood flow rate was 40 ml/min with the roller pump. The activated clotting time and the hematocrit level were adjusted to 170 +/- 10 s with heparin and 35 +/- 5% with a phosphate buffered sulfate solution, respectively. As a result, the backscattered light intensity from the low shear rate region gradually diminished with time (p < 0.05; 0-5 minutes vs. more than 10-15 minutes), whereas that from the high shear rate region remained fairly constant. The experimental finding supported the hypothesis that the decrease of the backscattered light intensity and diminished spectral power density were caused by the aggregation or clotting erythrocytes. In conclusion, we found that the measurement of the average backscattered light intensity level together with FFT analysis of the backscattered light from the flowing blood is a valuable approach in detecting the onset of thrombus formation.

Animals↗

Magnetically suspended centrifugal blood pump with a radial magnetic driver.

A new magnetic bearing has been designed to achieve a low electronic power requirement and high stiffness. The magnetic bearing consists of 1) radial passive forces between the permanent magnet ring mounted inside the impeller rotor and the electromagnet core materials in the pump casing and 2) radial active forces generated by the electromagnets using the two gap sensor signals. The magnetic bearing was assembled into a centrifugal rotary blood pump (CRBP) driven with a radial, magnetic coupled driver. The impeller vane shape was designed based upon the computational fluid dynamic simulation. The diameter and height of the CRBP were 75 mm and 50 mm, respectively. The magnetic bearing system required the power of 1.0-1.4 W. The radial impeller movement was controlled to within +/- 10 microm. High stiffness in the noncontrolled axes, Z, phi, and theta, was obtained by the passive magnetic forces. The pump flow of 5 L/min against 100 mm Hg head pressure was obtained at 1,800 rpm with the electrical to hydraulic efficiency being greater than 15%. The Normalized Index of Hemolysis (NIH) of the magnetic bearing CRBP was one fifth of the BioPump BP-80 and one half of the NIKKISO HPM-15 after 4 hours. The newly designed magnetic bearing with two degrees of freedom control in combination with optimized impeller vane was successful in achieving an excellent hemolytic performance in comparison with the clinical centrifugal blood pumps.

Biomedical Engineering↗

Feasibility of a miniature centrifugal rotary blood pump for low-flow circulation in children and infants.

In this study, a seal-less, tiny centrifugal rotary blood pump was designed for low-flow circulatory support in children and infants. The design was targeted to yield a compact and priming volume of 5 ml with a flow rate of 0.5-4 l/min against a head pressure of 40-100 mm Hg. To meet the design requirements, the first prototype had an impeller diameter of 30 mm with six straight vanes. The impeller was supported with a needle-type hydrodynamic bearing and was driven with a six-pole radial magnetic driver. The external pump dimensions included a pump head height of 20 mm, diameter of 49 mm, and priming volume of 5 ml. The weight was 150 g, including the motor driver. In the mock circulatory loop, using fresh porcine blood, the pump yielded a flow of 0.5-4.0 l/min against a head pressure of 40-100 mm Hg at a rotational speed of 1800-4000 rpm using 1/4" inflow and outflow conduits. The maximum flow and head pressure of 5.25 l/min and 244 mm Hg, respectively, were obtained at a rotational speed of 4400 rpm. The maximum electrical-to-hydraulic efficiency occurred at a flow rate of 1.5-3.5 l/min and at a rotational speed of 2000-4400 rpm. The normalized index of hemolysis, which was evaluated using fresh porcine blood, was 0.0076 g/100 l with the impeller in the down-mode and a bearing clearance of 0.1 mm. Further refinement in the bearing and magnetic coupler are required to improve the hemolytic performance of the pump. The durability of the needle-type hydrodynamic bearing and antithrombotic performance of the pump will be performed before clinical applications. The tiny centrifugal blood pump meets the flow requirements necessary to support the circulation of pediatric patients.

Centrifugation↗