PubMed Health⌕ Search

Biomedical subjects

Sven Zühlsdorff

Publications and source records attributed to Sven Zühlsdorff.

3 recordsLinked to original sources

Dynamic coil selection for real-time imaging in interventional MRI.

MR-guided intravascular interventions require image update rates of up to 10 images per second, which can be achieved using parallel imaging. However, parallel imaging requires many coil elements, which increases reconstruction times and thus compromises real-time image reconstruction. In this study a dynamic coil selection (DCS) algorithm is presented that selects a subset of receive coils to reduce image reconstruction times. The center-of-sensitivity coordinates and the relative signal intensities are determined for each coil in a prescan. During the intervention m coils are selected for reconstruction using a coil ranking based on the distance to the current slice or catheter position. In a phantom experiment for m = 6, an optimal signal-to-background ratio (SBR) was achieved and foldover artifacts were avoided. In three animal experiments involving catheter manipulation in the aorta and the right heart chamber, the anatomy was successfully visualized at frame rates of about 5 Hz using active catheter tracking.

Algorithms↗

MR-guided intravascular procedures: real-time parameter control and automated slice positioning with active tracking coils.

PURPOSE: To implement and optimize a real-time pulse sequence and user interface to perform intravascular interventions using active catheter tracking. MATERIALS AND METHODS: In magnetic resonance (MR)-guided interventions, small radio-frequency coils can be used to rapidly determine the device position (active tracking). In this work, active catheter tracking was combined with a dedicated real-time pulse sequence and user interface. The pulse sequence offered the imaging contrasts fast low angle shot (FLASH), true Fast imaging with steady state precession (TrueFISP), and projection MR digital subtraction angiography (MR-DSA), which could be selected by the radiologist from within the scanner room at any time during the intervention. Automatic slice positioning was added to the real-time pulse sequence so that the location of the tracking coils defined the image slice position and orientation. The technique was assessed in phantoms and animal experiments. RESULTS: At a reaction time of 24 msec and a frame rate of three images per second, the movement of an active intravascular catheter could be monitored in the aorta and the renal arteries of a pig. With interactive contrast and orientation changes, the renal vasculature could be assessed by a fully MR-guided catheterization in less than 10 minutes. CONCLUSION: With carefully designed active catheters, a dedicated user interface, and an optimized pulse sequence intravascular interventions can successfully be performed by a single operator from within the MR scanner room.

Animals↗

[Automatic slice tracking in interventional magnetic resonance imaging].

Magnetic resonance imaging (MRI) is ideally suited to monitor minimally invasive operations with catheters or needles, since it offers both a superior soft-tissue contrast and the possibility to perform functional tests. In the present study, small radio-frequency coils were attached to the instruments in order to localize the MR-invisible instruments. The implementation of active instrument tracking is described on the basis of the example of active catheter tracking. In this case, the current position information of the instrument is used to automatically position the MRI slice at the catheter location. In combination with a user interface, the interventional radiologist is offered the possibility to perform vascular interventions from within the MR scanner room. At image update rates of approximately 3 Hz, tracking and placement of catheters in vascular structures are possible with interactive switching of slice orientation and image contrast. In an animal model, the technique was successfully used to selectively visualize the abdominal vessels and their branches under MRI guidance.

Animals↗