Toward time-resolved soft X-ray microscopy using pulsed fs-high-harmonic radiation.

Coherent soft X-ray sources open the way to new capabilities in high-resolution imaging, site- and element-specific spectroscopy and biomicroscopy. In this paper we demonstrate imaging with a table-top soft X-ray microscope. By combining a laser driven high-harmonic light source, optimized for having the maximum brightness at around 100 eV, a pair of multilayer mirrors to select a narrow spectral band and acting simultaneously as a condenser and a Fresnel zone plate as microscope objective, we were able to resolve 200 nm structures of a diatom sample. Further, the pulsed nature of our X-ray source offers the possibility of time-resolved spectromicroscopy with a temporal resolution in the order of a few femtoseconds.

Animals↗

Compact water-window transmission X-ray microscopy.

We demonstrate sub-100 nm resolution water-window soft X-ray full-field transmission microscopy with a compact system. The microscope operates at lambda = 3.37 nm and is based on a 100 Hz table-top regenerative debris-free droplet-target laser-plasma X-ray source in combination with normal-incidence multilayer condenser optics for sample illumination. High-spatial-resolution imaging is performed with a 7.3% efficiency nickel zone plate and a 1024 x 1024 pixel CCD detector. Images of dry test samples are recorded with exposure times of a few minutes and show features smaller than 60 nm.

Diatoms↗