Computerized planning for multiprobe cryosurgery using a force-field analogy.

Cryosurgery is the destruction of undesired biological tissues by freezing. For internal organs, multiple cryoprobes are inserted into the tissue with the goal of maximizing cryoinjury within a predefined target region, while minimizing cryoinjury to the surrounding tissues. The objective of this study is to develop a computerized planning tool to determine the best locations to insert the cryoprobes, based on bioheat transfer simulations. This tool is general and suitable for all available cooling techniques and hardware. The planning procedure employs a novel iterative optimization technique based on a force-field analogy. In each iteration, a single transient bioheat transfer simulation of the cryoprocedure is computed. At the end of the simulation, regions of tissue that would have undesired temperatures apply "forces" to the cryoprobes directly moving them to better locations. This method is more efficient than traditional numerical optimization techniques, because it requires significantly fewer bioheat transfer simulations for each iteration of planning. For demonstration purposes, 2D examples on cross sections typical of prostate cryosurgery are given.

Algorithms↗

Computerized planning of cryosurgery using cryoprobes and cryoheaters.

In a typical minimally invasive cryoprocedure, multiple cryoprobes are inserted into the tissue with the goal of maximizing cryoinjury within a predefined target region, while minimizing cryoinjury to the surrounding tissues. A temperature-controlled electrical heater has been developed recently by this research team, in order to assist in limiting the cryoinjury to the target region. The new device has been termed a 'cryoheater,' and it can work with any cryosurgical cooling technique. A prototype computerized planning tool has been presented recently by this research team, which helps to determine the best locations in which to insert the cryoprobes. This prototype was designed for cryoprobes only. The planning procedure utilized a novel iterative optimization technique, based on a force-field analogy. The combination of cryoheaters with computerized planning is the subject matter of this report. The current report includes a review of cryoheater development, and presents an improved cryosurgery planning tool which incorporates cryoheaters.

Computer Simulation↗

Cryoheater as a means of cryosurgery control.

A new concept for cryosurgery control is presented in this paper, a concept which has the potential to dramatically change the outcome of cryosurgery. Unlike other cryosurgery control techniques, which are based on controlling the thermal performance of the cryoprobe, this new concept is based on heating the treated tissue as a means of shaping the frozen region. The new controlling heater is termed a 'cryoheater'. The cryoheater is a complementary device to the cryoprobe and can work with any cryosurgery cooling technique. In the current pilot study, the new cryoheater is demonstrated in a gelatin solution and, using heat transfer simulations, it is studied in simulated cases of prostate cryosurgery. It is suggested that cryosurgery planning tools and optimization techniques must be developed before the concept of the cryoheater can be applied in its full capacity.

Body Temperature↗

The thermal effect of urethral warming during cryosurgery.

The heating effect of urethral warming during cryosurgery has been investigated theoretically, via heat transfer simulations. Two warmer configurations have been considered: (i). the clinically available urethral warmer, which has a configuration of a counter flow fluid heat exchanger; (ii). a newly designed urethral warmer, based on a temperature controlled electrical heater, termed a "cryoheater". A dramatic effect of thermal resistance to heat transfer through the heat exchanger wall has been identified, which is absent in the cryoheater. It follows that the cryoheater is expected to be more efficient in generating an unfrozen region around the urethra. It is shown that the conventional heat exchanger may fail to prevent freezing around the urethra in a significant number of prostate cases, depending on the layout of cryoprobes around the urethra. On the other hand, clinical reports exist which suggest that the heat exchanger improves in many cases the outcome of cryosurgery, in terms of long term complications. It is speculated in the current report that the cryoheater can further improve the outcome of cryosurgery, by providing protection from freezing in a wider range of cases. It is suggested that a future study be conducted to examine the correlation between the layout of cryoprobes and surgical outcome.

Cryosurgery↗