PubMed Health⌕ Search

Biomedical subjects

Trefford L Simpson

Publications and source records attributed to Trefford L Simpson.

15 recordsLinked to original sources

Overnight corneal swelling with silicone hydrogel contact lenses with high oxygen transmissibility.

PURPOSE: To compare central corneal swelling after 8 hours of sleep in eyes wearing 2 different silicone hydrogel lenses with high oxygen transmissibility. METHODS: Twenty neophyte subjects were randomly assigned to wear a comfilcon A lens (CooperVision Inc) in one eye and a lotrafilcon A lens (CIBA Vision) in the contralateral eye for an 8-hour overnight period. The study was repeated with another 20 neophyte subjects wearing comfilcon A in one eye only and no lens in the contralateral control eye. Central corneal thickness was measured with optical pachometry before lens insertion, immediately after lens removal on waking, at 20 and 40 minutes, 1 hour, 2 hours and 3 hours. RESULTS: There was no significant difference between the amount of overnight central corneal swelling induced by the 2 lenses (4.1% +/- 1.9% with comfilcon A vs. 4.0% +/- 1.7% with lotrafilcon A; P>0.05). The swelling of the comfilcon A lens-wearing eyes was significantly higher than that found in the non-lens-wearing contralateral eyes (4.5% +/- 2.1% vs. 3.0% +/- 1.7%; P<0.05). CONCLUSIONS: The same amount of overnight corneal swelling was induced by the 2 study lenses. The higher overnight swelling observed with the comfilcon A lenses compared to the non-lens-wearing eyes is similar to that reported in the literature for studies using lotrafilcon A lenses.

Adult↗

The repeatability of corneal and corneal epithelial thickness measurements using optical coherence tomography.

PURPOSE: The purpose of this study is to examine within and between session repeatability of clinical optical coherence tomography (OCT) imaging for anterior segment morphometry. METHODS: Images of the corneal apex of each eye in 18 subjects were obtained using a Humphrey Zeiss OCT imager. Subjects viewed a target positioned to ensure that scans were orthogonal to the ocular surface and each image, consisting of 100 adjacent sagittal scans, analyzed using custom software. Repeatability data were analyzed using intraclass correlation coefficients (ICCs), correlation coefficient of concordance (CCC, perfect test-retest agreement ICC or CCC = 1.0), and coefficients of repeatability (COR, 95% confidence interval of test-retest differences). To account for each eye, the multivariate repeatability statistic Iota was estimated. RESULTS: Mean central corneal and epithelial thickness of 32 eyes (OD and OS combined) is 536 +/- 26 microm (standard deviation [SD]) and 52 +/- 3 microm (SD) with 5th and 95th percentile thicknesses of 507 and 591 microm for central cornea and 48 and 57 microm for central epithelial. Worst case within session repeatability was defined as repeatability between images with greatest differences in mean thickness within a session. Corneal thickness worst case ICC was 0.95 and COR was +/- 9.98 microm. Epithelium worst case ICC was 0.36, CCC was 0.12, and COR was +/- 11.11 microm. First image between session corneal thickness had an ICC = 0.98 and a COR = 10.83 microm, whereas epithelium ICC = 0.38, CCC = 0.37, and COR was +/- 12.84 microm. When we compared the average of the first three tests with the first three retest images, corneal ICC was 0.98 and COR was +/- 10.64 microm and epithelium ICC = 0.73, CCC = 0.72, and COR was +/- 6.53 microm. Iota (multivariate repeatability, using eye as a factor) for the cornea was at least 0.96 (worst case) and increased to at least 0.98 when within-session image data were averaged. Iota for epithelium measures ranged from 0.29 when first images were compared with 0.57 when within-session image data were averaged. CONCLUSIONS: There is very good repeatability of corneal thickness measurement using OCT; even the worst case measurements are similar between sessions. On the other hand, this is not the case for epithelium measurements, and if multiple images within a session are acquired, the worst case results demonstrate how important it is to optimize each OCT scan and also average multiple scans to maximize intersession repeatability.

Adolescent↗

Novel pachometry calibration.

PURPOSE: The purpose of this study was to develop a simple method for cross-calibrating instruments that measure corneal thickness. METHODS: Fourteen rigid lenses of different thicknesses were manufactured using a material with refractive index of 1.376. Center thickness of the lenses (CT) was measured using a computerized optical pachometer (OP), two optical coherence tomographers (OCTs), and a confocal microscope (CM). Accuracy of measurements was compared between the four instruments. RESULTS: Before calibrating the machines, there was a significant effect of the measurement device (p < 0.05). The differences between instruments were eliminated (p > 0.05) after applying calibration equations for each device. In addition, after each instrument was calibrated with lenses of 1.376 refractive index, there was no significant difference (p > 0.05) between measured values of lens center thickness by OP, each OCT, CM, and the physical center thickness of the lenses. CONCLUSIONS: Using calibration lenses with the same refractive index as the cornea (1.376) allows rapid and simple calibration of the pachometers so that corneal thickness measurements from different devices can be used interchangeably.

Calibration↗

Comparison of human central cornea and limbus in vivo using optical coherence tomography.

PURPOSE: The purpose of this study was to compare central corneal and limbal total and epithelial thickness using a commercially available optical coherence tomographer. METHODS: A Humphrey-Zeiss Optical Coherence Tomographer (OCT [Carl Zeiss, Meditec, Dublin, CA]) was used to obtain corneal images from 10 subjects. Central corneal and limbal total and epithelial thickness of both eyes were measured using the OCT. Each OCT image comprised 100 measurements, 10 nasal, 10 central, and 10 temporal measurements from each image were analysed. RESULTS: The central corneal and epithelial thickness of the right and the left eyes were 507.9 +/- 35.8 microm, 58.4 +/- 2.5 microm, 506.9 +/- 37.4 microm, and 58.5 +/- 2.5 microm, respectively. There were no differences between eyes (p > 0.05). The nasal and temporal limbal total and epithelial thickness of the right and left eyes were 703.8 +/- 32.1 microm, 704.9 +/- 31.0 microm, 76.8 +/- 3.5 microm, 77.9 +/- 2.9 microm, 704.4 +/- 31.8 microm, 706.3 +/- 32.5 microm, 77.5 +/- 2.8 microm, and 77.8 +/- 2.5 microm, respectively. There were no differences between the nasal and temporal total and epithelial thickness of both eyes (p > 0.05). However, there was a statistical difference between the central corneal and limbal total and epithelial thickness (both p < 0.05). CONCLUSIONS: Central cornea and limbus are measurably different using OCT. Central cornea is thinner than limbus for both total thickness and epithelial thickness. There is no difference between eyes of central corneal and limbal total and epithelial thickness. Optical Coherence Tomography is a useful instrument for in vivo human limbal morphometry.

Adult↗

The inhibitory interaction between human corneal and conjunctival sensory channels.

PURPOSE: To explore human corneal and conjunctival sensory channels at suprathreshold level. METHODS: Ten healthy human subjects participated in the study. The Belmonte pneumatic esthesiometer was used to apply mechanical and chemical stimuli to the central cornea and temporal conjunctiva of the left eye. Stimuli were applied in a paired and unpaired way for conjunctival stimulation. A 100-point visual analog scale (VAS) was used to rate the intensity of the stimulus. RESULTS: The magnitudes of the sensation evoked from the conjunctiva were different when using different methods for presenting stimuli to the ocular surface. When stimuli were applied to the conjunctiva alone, the magnitude of the sensation was stronger than when the stimuli were applied in pairs to the cornea and conjunctiva for both mechanical (P = 0.04) and chemical (P = 0.02) stimulation. CONCLUSIONS: The relatively strong discomfort evoked from the cornea appears to suppress partially the relatively weaker conjunctival stimulation. This manifested as the conjunctival sensory transducer function being shallower (less intense sensation) when immediately preceded by corneal stimulation than when the conjunctival sensory transducer functions were measured alone (unpaired). The underlying mechanism could be adaptation or some other inhibitory effect, such as diffuse noxious inhibitory control. At some level therefore, corneal and conjunctival sensory channels are not independent.

Adult↗

Frequency doubling illusion: detection vs. form resolution.

PURPOSE: To investigate the difference between detection and form resolution thresholds for the frequency doubling (FD) illusion. METHOD: The right eyes of 20 normal healthy subjects (10 female; age range, 18 to 30 years; mean age, 24.05 years; SD, 1.39) were examined. Vertically oriented FD stimuli were presented at fixation and 10 degrees nasally using a dual temporal alternate forced choice method of constant stimuli (2AFC MOCS) to estimate the thresholds for detection and form resolution. Additional strategies for threshold estimation (yes/no, modified rapid estimation by binary search) were used to determine the detection threshold. The effect of subject instruction on the FD threshold was also examined. The test-retest characteristics were investigated by determining the coefficient of repeatability. RESULTS: Detection thresholds using 2AFC MOCS were 0.86 +/- 0.20 (+/-1 SE) Michelson contrast percentage (MC%) at fixation and 0.84 +/- 0.21 at 10 degrees. Form resolution thresholds using 2AFC MOCS were 1.08 +/- 0.23 MC% at fixation and 1.04 +/- 0.16 at 10 degrees . These thresholds were found to be significantly different at fixation (p = 0.004) and 10 degrees (p = 0.005). No difference was found between threshold estimation strategies, but subject instruction had a significant effect (fixation: 2.41 +/- 0.52 MC%, p < 0.001; 10 degrees : 2.19 +/- 0.48 MC%, p < 0.001). CONCLUSIONS: Detection thresholds were significantly lower than form resolution thresholds for the FD stimulus. This result is in agreement with classic FD studies and illustrates that the perception of flicker precedes the perception of the FD illusion at threshold.

Adult↗

Contact lens-induced corneal swelling and surface changes measured with the Orbscan II corneal topographer.

PURPOSE: The purpose of this study was to measure central and topographical corneal swelling in response to contact lens wear and eye closure, to determine whether the swelling induced by soft and polymethyl methacrylate (PMMA) lenses is different, and to determine whether the anterior and/or posterior corneal shape alters with corneal swelling. METHODS: An Orbscan II corneal topographer was used to measure corneal swelling and the shape of the anterior and posterior corneal surfaces of 16 neophytes before and after wearing soft and PMMA contact lenses with near-zero oxygen transmissibility. The lens-wearing eye was patched for 3 h. RESULTS: The mean 15.1% +/- 3.8% (+/- SD) central corneal swelling with soft lenses was significantly >12.6% +/- 4.1% with PMMA lenses. Topographical corneal swelling was significantly greater with soft vs. PMMA lenses. However, the difference between central and peripheral corneal swelling was much greater with PMMA lenses. With both lenses, the cornea swelled significantly more in the center than the periphery. The anterior best-fit sphere radius remained unchanged in response to soft lenses (0.00 +/- 0.04 mm) and steepened slightly but significantly with PMMA lenses (-0.04 +/- 0.05 mm). The posterior best-fit sphere radius flattened significantly with both lenses (0.12 +/- 0.07 mm with soft and 0.14 +/- 0.08 mm with PMMA lenses). CONCLUSIONS: Corneal swelling (greater centrally than peripherally) flattens the posterior surface of the cornea and is independent of lens type. Although the anterior best-fit sphere radius steepened with PMMA, the magnitude is probably clinically unimportant. Both lens types produced greater central vs. peripheral corneal swelling. However, the soft lens induced significantly greater overall swelling than PMMA. Because their oxygen transmission was the same, these results suggest that there is lateral diffusion of oxygen from the peripheral area of the cornea (that is not covered by the lens) toward the center.

Adult↗

Objective measurements of corneal light-backscatter during corneal swelling, by optical coherence tomography.

PURPOSE: To demonstrate that corneal light-backscatter can be measured objectively during corneal swelling by optical coherence tomography (OCT). METHODS: One eye (randomly selected) of 20 non-contact-lens wearers (10 men and 10 women; mean age, 35.6 +/- 9.6 years) was patched during 3 hours of soft contact lens (SCL) wear. The contralateral eye acted as the control. Central corneal images were captured before and after SCL wear at 20-minute intervals over 100 minutes using optical coherence tomography (OCT) to obtain corneal thickness and light-backscatter profiles. OCT backscattered light of the epithelial layer (decided by the thickness measurements) and 10 equally divided layers of the remaining cornea were analyzed with a custom software program. Two baseline measurements were taken at different visits before lens wear to test the repeatability of light-backscatter measurements. RESULTS: From two baseline measurements, repeated measurements showed good repeatability of normalized backscatter results. Immediately after contact lens removal, total central corneal thickness increased significantly by 13.8% +/- 2.3% (mean +/- SD) compared with baseline (P = 0.0001, paired t-test) and then decreased during the deswelling course. Corneal backscattered light changed significantly (repeated-measures ANOVA [Re-ANOVA]: F((50, 950)) = 2.22, P = 0.0001) after lens wear, and a significant increase in backscatter was found in the epithelial layer (36.4%) and the most posterior corneal layer (35.6%) immediately after lens removal (post hoc test, P = 0.005). There was a strong correlation (r = 0.9375, P < 0.05) between the change in backscatter and corneal swelling during the deswelling period. The backscatter recovery rate was approximately the same for both epithelial and posterior layers after lens removal. CONCLUSIONS: Light-backscattering analysis with OCT seems to be a promising and repeatable method of objectively measuring corneal backscatter. This study has demonstrated that corneal backscattered light increased in the anterior and posterior layers of the cornea during corneal swelling induced by contact lens wear and eye closure.

Adult↗

Characteristics of human corneal psychophysical channels.

PURPOSE: To characterize human corneal psychophysical channels. METHODS: Twenty subjects participated in this study. A Belmonte pneumatic esthesiometer was used to deliver stimuli, and the ascending method of limits and the method of constant stimuli were used to estimate thresholds. Sensation was characterized for different stimuli. Corneal mechanical and chemical thresholds were measured at different temperatures. RESULTS: The qualities of the sensations induced by stimuli with different temperatures were different, and the corresponding detection thresholds of the pneumatic stimuli at four temperatures gradually increased (repeated measures ANOVA (F(3,12) = 10.326, P = 0.000). There were no temperature effects on chemical thresholds (repeated measures ANOVA F(3,12) = 0.235, P = 0.870) or mechanical discomfort thresholds from 20 degrees C to 50 degrees C (paired t14 = -0.233, P = 0.818). There were strong interactions when chemical and mechanical stimuli were added. Chemical thresholds were progressively lower when the flow rate increased and mechanical thresholds went down as the percentage of added CO2 increased (repeated measures ANOVA F(3, 12) = 6.407, P = 0.007, F(4, 16) = 19.904, P = 0.000). CONCLUSIONS: This study suggests that humans sense corneal non-noxious cold and noxious mechanical and chemical stimuli, and that the sensitivity of some submodalities can be modulated by others. There are at least five psychophysical channels (non-noxious cold, noxious mechanical, noxious chemical-H+, noxious heat, and itching) processing corneal sensory information. Both decreased corneal chemical thresholds at high flow rates and decreased mechanical thresholds with an added chemical stimulation demonstrate that corneal psychophysical channels are not independent.

Adult↗

Nociceptive sensation and sensitivity evoked from human cornea and conjunctiva stimulated by CO2.

PURPOSE: To compare sensation and sensitivity evoked from human cornea and conjunctiva stimulated by CO2. METHODS: Twenty healthy participants were recruited for the study. Central corneal and temporal conjunctival chemical sensation and sensitivity of only one eye of each subject were evaluated. Air mixed with different concentrations of CO(2) was delivered by a modified Belmonte pneumatic esthesiometer. The ascending method of limits was used to determine the sensitivity and subjects were required to characterize the sensation at threshold. RESULTS: The sensations evoked by CO(2) in the cornea and conjunctiva were stinging or burning. The sensation evoked by mechanical stimulation was that of irritation. The corneal and conjunctival chemical thresholds were 31% +/- 2% and 54% +/- 5% CO(2) (mean +/- SE), respectively. The corneal and conjunctival mechanical thresholds were 80 +/- 6 and 140 +/- 10 mL/min (mean +/- SE), respectively. The corneal sensitivity was significantly higher for both mechanical and chemical stimuli (P < 0.05). CONCLUSIONS: The results suggest that CO(2) stimulates similar corneal and conjunctival nociceptors in that the interpretations were the same (i.e., nociceptive). The central cornea had a higher sensitivity to CO(2) than the temporal conjunctiva, which may reflect a different peripheral innervation, such as different nerve density or different receptor characteristics. Sensations evoked by mechanical and chemical stimulation were different, which suggests that at the peripheral level, the two modalities stimulate two different kinds of molecular receptors or channels and that this information is somehow retained within the nociceptive system.

Adult↗

Topographical thickness of the epithelium and total cornea after hydrogel and PMMA contact lens wear with eye closure.

PURPOSE: To determine changes in topographical thickness of the epithelium and total cornea after hydrogel (2-hydroxyethyl methacrylate; HEMA or soft lens) and PMMA rigid contact lens wear with eyes closed, as measured by optical coherence tomography (OCT). METHODS: Epithelial and total corneal thickness in 18 neophyte eyes was measured with OCT at intervals of 10 degrees across a 10-mm zone of the horizontal meridian of the cornea, before and after 3 hours of soft and rigid contact lens wear with the eye closed. These measurements were repeated 20 minutes after removal of the lenses. RESULTS: Lens type, time, and location were found to be significant main influences (P < 0.0001) on corneal swelling in patched eyes, by three-way ANOVA, and there was a significant three-way interaction among lens type, time, and location (F((16,272)) = 1.78, P = 0.033). However, there was no significant main effect and interaction of epithelial thickness (F((16, 272)) = 0.33, P = 0.99). Immediately after removal of the lenses, total corneal thickness in the horizontal meridian was significantly greater with both soft and PMMA lenses (P < 0.001) at each location with each lens, compared with the baseline measurements. With both lenses, the increase in actual thickness and percentage of corneal swelling at the center was greater than at each peripheral point (excluding the first 10 degrees points; P < 0.005). HEMA lenses caused greater corneal swelling than the PMMA lenses at each location immediately after removal of the lenses (P < 0.005). CONCLUSIONS: This study shows that corneal swelling is dependent on lens type and corneal location when eyes are closed, but epithelial thickness across the horizontal corneal meridian does not change during lens wear with eyes closed. OCT is an efficient method of measuring topographical corneal and epithelial thickness in response to contact lens wear.

Adult↗

Precorneal and pre- and postlens tear film thickness measured indirectly with optical coherence tomography.

PURPOSE: To demonstrate the feasibility of indirectly measuring the precorneal tear film thickness and pre- and postlens tear film (PLTF) thickness using optical coherence tomography (OCT). METHODS: Central corneal thickness (C(1)) which includes the tear film (T) of both eyes of 40 non-contact lens wearers was measured using OCT after calibration. The mean age of the 40 subjects was 31.2 +/- 9.3 years with a mean horizontal K-reading of 7.87 mm. Rigid contact lenses with base curves 0.3- to 0.5-mm steeper than the flattest K of the eye were fitted to measure real corneal thickness (C(2)), independently of the postlens tear film. T was calculated by T = C(1) - C(2). To measure pre- and postlens tear film thickness, Focus Night & Day and Acuvue lenses (Vistakon, Johnson & Johnson Vision Care, Jacksonville, FL) were fitted on both eyes. Central soft lens thickness (L(1)), which includes the prelens tear film (P), was measured by OCT in situ and in saline in a wet cell (L(2)). P was calculated by P = L(1) - L(2). Thickness of the central cornea plus the postlens tear film (C(3)) was measured during lens wearing. Postlens tear film (PLTF) was calculated by PLTF = C(3) - C(2). RESULTS. The mean +/- SD precorneal tear film thickness was 3.3 +/- 1.5 microm (range, 0-6.9) before lens insertion and 4.7 +/- 2.3 microm (range, 0.7-11.0) after lens fitting, which was significantly thicker (paired t-test: P < 0.01). The prelens tear film thickness was 3.9 +/- 2.6 and 3.6 +/- 2.1 microm (mean +/- SD; paired t-test: P = 0.52) and the postlens tear film thickness was 4.5 +/- 2.3 and 4.7 +/- 3.1 microm (paired t-test: P = 0.08) on and under Focus Night & Day and Acuvue lenses, respectively. Post hoc tests showed that precorneal (baseline) and prelens tear films were equivalent, and each was different (thinner; Tukey honestly significant difference P < 0.05) from the postlens tear film. CONCLUSIONS: OCT can noninvasively measure the thickness of the precorneal and prelens tear film as well as the postlens tear film. The thickness of the normal precorneal tear film is approximately 3 micro m and becomes thicker after lens fitting. The postlens tear film is thicker than the precorneal and prelens tear films with soft contact lenses. The thickness of both pre- and postlens tear films appears to be independent of the investigated lens types.

Adult↗

Topographical thickness of the epithelium and total cornea after overnight wear of reverse-geometry rigid contact lenses for myopia reduction.

PURPOSE: To investigate topographical thickness changes of the epithelium and total cornea measured with optical coherence tomography (OCT) after overnight wear of rigid gas-permeable lenses. METHODS: Reverse geometry design CRT (Dk=100) rigid (test) lenses (Paragon Vision Sciences, Mesa, AZ) were randomly fitted on one eye of each of 20 neophyte subjects (mean age, 24.6 +/- 2.7 years) and the other eye was fitted with an alignment tricurve rigid lens of the same material (control). Epithelial and total corneal thickness was measured at intervals of 10 degrees across a 10-mm zone of the horizontal meridian of the cornea, before and after overnight wear. Refractive error was measured with an autorefractor. These measurements were repeated 20 and 60 minutes and 3, 6, and 12 hours after lens removal. RESULTS: After one night of lens wear, myopia decreased in the test eye by 1.18 +/- 0.81 D, which was significantly different from baseline (P<0.001). No significant change was found in the control eye. Twelve hours after removal, two thirds of the myopic reduction was still present. Different topographical swelling patterns were induced by the two lens designs, with greatest swelling occurring in the center with the control lens and in the midperiphery with the test lenses (polynomial regression: P<0.005). Significantly greater central corneal swelling was found with the control lens than the test lens (6.9% +/- 3.1% vs. 4.9% +/- 2.0%, respectively, P=0.006). The effect on epithelial thickness differed between lenses and depended on both position and time (F(48,912)=2.3; P=0.000). Immediately after removal of the test lens, the central epithelium was 5.1% +/- 4.5% thinner than baseline, and all other locations (P<0.005 post hoc tests) and the epithelium in the midperiphery showed significant thickening (1.9% on the temporal side and 2.4% on the nasal side, both P<0.006 compared with the baseline). There were no significant changes in epithelial thickness with the control lens during the study period (post hoc tests: P>0.05). CONCLUSIONS: The optical coherence tomograph is a sensitive instrument that can detect small changes in epithelial and corneal thickness across the entire cornea. Topographical thickness changes of the epithelium and total cornea induced by one night of reverse-geometry lens wear appear to be associated with the decrease of myopia.

Adult↗

The measurement of corneal epithelial thickness in response to hypoxia using optical coherence tomography.

PURPOSE: To determine if corneal epithelial thickness increases in association with corneal edema induced by wearing soft contact lenses during eye closure. DESIGN: Experimental study. METHODS: One eye (randomly selected) of twenty noncontact lens wearers (10 males and 10 females, age 35.6 +/- 9.6 years) was patched during 3 hours of soft contact lens (SCL) wear and the contralateral eye acted as a control. Corneal and epithelial thickness of both eyes was measured before and after SCL wear using optical coherence tomography (OCT). RESULTS: Immediately after contact lens removal, total corneal thickness was increased significantly by 13.8 +/- 2.3% (mean +/- SD) compared with baseline (P <.0001, paired t test) and after 100 minutes was still 4.5 +/- 2.3% thicker than baseline (P <.0001, paired t test). The control eyes showed no change in total corneal thickness (P >.05, paired t test). Immediately after contact lens removal, corneal epithelial thickness was increased by 1.7 +/- 4.8%, but this change was not statistically significant (P >.05, paired t test). Following contact lens removal, epithelial thickness changed significantly (Repeated measure analysis of variance [Re-ANOVA]: F((7,133)) = 4.91, p(H-F) < 0.001) over the next 100 minutes with thinning recorded at 60, 80, and 100 minutes (P <.05, paired t test). There was no significant change over time in epithelial thickness of the control eyes (Re-ANOVA: F(4, 76) = 0.91, p(H-F) = 0.464). CONCLUSION: OCT demonstrated that corneal epithelial thickness does not increase in response to hypoxia from SCL wear and eye closure, in contrast to a significant increase in total corneal thickness.

Adult↗

Relation between optical coherence tomography and optical pachymetry measurements of corneal swelling induced by hypoxia.

PURPOSE: To determine the relation between optical coherence tomography (OCT) and optical pachymetry (OP) measurements of corneal swelling induced by hypoxia. DESIGN: Experimental study. METHODS: One randomly selected eye of 20 noncontact lens wearers (10 males and 10 females, age 35.6 +/- 9.6 years) was patched during 3 hours of soft contact lens (SCL) wear while the contralateral eye acted as control. Central corneal thickness of both eyes was measured before and after SCL wear using OCT and OP in randomized order. RESULTS: Baseline central corneal thickness was 523.6 +/- 33.0 microm (mean +/- standard deviation [SD]) measured with OCT and 490.6 +/- 25.5 microm with OP. Immediately after contact lens removal, corneal thickness measured with OCT increased by 13.8 +/- 2.3% compared with 12.1 +/- 1.8% (paired t test: P <.001) measured with OP. Thereafter, corneal thickness decreased at the rate of 5.6% per hour for OCT and 5.4% per hour for OP. The difference in thickness between instruments before lens insertion, which was 33 microm compared with the difference after lens removal (edematous cornea), which ranged from 46 to 41 microm. The difference between instruments decreased during the corneal deswelling period after lens removal. The correlation coefficient between OCT and OP was 0.914 before lens insertion and 0.932 after lens removal. CONCLUSION: This study has demonstrated the difference of corneal thickness measured with OCT and OP. Although both instruments are correlated highly in all conditions tested, OCT may overestimate corneal thickness in normal and edematous corneas.

Adult↗