A robust multi-syringe system for process flow analysis. Part 3. Time based injection applied to the spectrophotometric determination of nickel(II) and iron speciation.

A new software-controlled time-based system for sample or reagent introduction in process flow injection analysis was developed. By using a multi-syringe burette coupled with one multi-port selection valve, the time-based injection of precise known volumes was accomplished. Characteristics and performance of the injection system were studied by injecting an indicator in a buffered carrier. Two multi-syringe time-based injection (MS-TBI) systems were implemented: first, the injection of a sample in a multiple-channel manifold where the sample would sequentially merge and react with different reagents, and second, the sequential injection of several solutions (sample and reagents) into a particular flowing stream. The first system was applied to the spectrophotometric determination of nickel(II) in diluted samples from the acidic nickel ore leaching process, by using ammonium citrate as carrier, a saturated solution of iodine as oxidizing agent and alkaline dimethylglyoxime as chromogenic reagent. The sampling frequency attained was 57 h-1. Determinations on process samples compared well at the 95% confidence level with the reference values obtained by ICP-OES. The second time-based injection system was applied to the speciation of iron. Total iron and iron(II) concentrations were separately and sequentially determined using 1,10-phenanthroline in acetic buffer medium as reagent. The developed manifold allowed the optional use of two different carrier solutions, containing or not containing ascorbic acid, for performing the separate determinations. Also, in the sequential procedure, plugs of reducing carrier were alternatively intercalated before the sample injections used for total iron determinations. Sampling frequencies of 68 injections per hour were routinely used. Accuracy was assessed by analyzing synthetic known mixtures of Fe(III) and Fe(II) standard solutions. Recoveries of 98-100.5% with a maximum relative standard deviation of 3.6% were found. Results obtained for various samples of fertilizers agreed well with those attained by the standard batch procedure.

Journal Article↗

Fast sequential injection determination of benzo[A]pyrene using variable angle fluorescence with on-line solid-phase extraction.

A methodology for the analysis of drinking water for one of the most potent carcinogenic agents known; benzo[a]pyrene (BaP), in the presence of other interfering PAHs is presented. The methodology described is based on the sequential injection analysis of the sample on to a microcolumn (containing 5 mg of C18) where extraction and preconcentration of BaP takes place, followed by elution of BaP with 1 ml of 1,4-dioxane and subsequent detection by using variable angle fluorescence. The advantages of the method include the small amount of stationary phase employed together with the possibility of re-using the phase in order to carry out a large number of injections without the need for column re-packing. Also noteworthy is the small volume of 1,4-dioxane used to elute the BaP retained on the column and the small sample volumes required (9-10 ml) for achieving detection limits at the ng l-1 level. Thus, a methodology for BaP determination is obtained which complies with the requirements of the 98/83/EC Directive which fixes a maximum admissible concentration for this pollutant in waters for public consumption of 10 ng l-1. The variable angle spectra obtained are further processed by means of the multiple linear regression technique. The detection limit for BaP is 2.5 ng l-1, and the linear range is between 7.5 and 280 ng l-1.

Autoanalysis↗

Simultaneous determination of phosphate and silicate in waste water by sequential injection analysis.

A sequential injection analysis system for the simultaneous determination of phosphate and silicate in waste water is proposed. The method is based on the formation of yellow vanadomolybdophosphate and molybdosilicate, respectively, in addition to the use of large sample volumes. The mutual interference between both analytes was eliminated by selection of the appropriate acidity and by sample segmentation with oxalic acid. The calibration graph for phosphate and silicate is linear up to 12 mg l-1 P and 30 mg l-1 Si, respectively. The detection limits are 0.2 mg l-1 P and 0.9 mg l-1 Si. The method provides a throughput of 23 samples h-1 with a relative standard deviation < 1.4% for phosphate and < 4% for silicate. The method was found to be suitable for the determination of these species in waste water samples.

Evaluation Studies as Topic↗

Multisyringe flow injection analysis of stable and radioactive strontium in samples of environmental interest.

A multisyringe flow injection (MSFIA) method for the determination of stable and radioactive strontium, using a solid phase resin (Sr-Resin), has been developed. Strontium concentrations are determined by atomic emission spectroscopy and by a low background proportional counter. The method has been applied to different samples (water, milk and soil) of environmental interest. The LLD of the stable and radioactive Sr were 10 microg/l and 0.01 Bq, respectively. The standard deviation of the separation procedure is 2% (n=10).

Animals↗

Sequential-injection procedure for determination of iodide in pharmaceutical and drinking water samples by catalytic reaction with spectrophotometric detection.

A novel sequential-injection system was developed for determination of iodide at very low concentrations by using a kinetic method. The method is based on the catalytic effect of iodide on the redox reaction between Ce4+ and As3+ first described by Sandell and Kolthoff. The calibration curve is constructed by measuring the decrease of Ce4+ absorbance versus iodide concentration with a delay time of 30 s. The detection limit is 1.5 microg/L, the working temperature is 45 degrees C, and the working range is 0.002-0.5 mg/L. Reasonable agreement was obtained when the method was applied to pharmaceutical and drinking water samples. The method has a sample throughput of approximately 15/h.

Arsenic↗