[Attempt at creating the 1st Soviet AB-1 biochemical autoanalyser].
Explore the source record for details and available documents.
Autoanalysis↗
Biomedical subjects
Publications and source records attributed to V E Deev.
Explore the source record for details and available documents.
An information processing system (IPS) is described for checking the parity of the sample address code, setting the scale factor of the measured value from 0.1 to 1.0 analog-digital conversion of this value with automatic selection of measuring subranges. The IPS performs printing of the sample address and the measurement results in red or black colour depending on the autoanalyzer operation mode and on the input signals. In addition, it sends control signals into the autoanalyzer to check its functioning. This IPS is intended for operation with DA240-4 discrete autoanalyzer.