Fluorescence molecular tomography in the presence of background fluorescence.

Fluorescence molecular tomography is an emerging imaging technique that resolves the bio-distribution of engineered fluorescent probes developed for in vivo reporting of specific cellular and sub-cellular targets. The method can detect fluorochromes in picomole amounts or less, imaged through entire animals, but the detection sensitivity and imaging performance drop in the presence of background, non-specific fluorescence. In this study, we carried out a theoretical and an experimental investigation on the effect of background fluorescence on the measured signal and on the tomographic reconstruction. We further examined the performance of three subtraction methods based on physical models of photon propagation, using experimental data on phantoms and small animals. We show that the data pre-processing with subtraction schemes can improve image quality and quantification when non-specific background florescence is present.

Algorithms↗

Inflammation in atherosclerosis: visualizing matrix metalloproteinase action in macrophages in vivo.

BACKGROUND: Matrix metalloproteinases (MMPs) in inflamed atherosclerotic plaques may contribute to extracellular matrix remodeling and the onset of acute thrombotic complications. METHODS AND RESULTS: To test the hypothesis that optical molecular imaging with the use of an activatable near-infrared fluorescence (NIRF) probe can detect enzymatic action of MMP in atherosclerotic plaques, we used a NIRF substrate for gelatinases (MMP-2/gelatinase-A and MMP-9/gelatinase-B) in apolipoprotein E-deficient (apoE-/-) mice that consumed a high-cholesterol diet for 12 weeks and age-matched apoE+/+ mice as control. The aortas of apoE-/- mice at 24 hours after probe yielded intense NIRF signals, as detected by NIRF reflectance ex vivo, compared with negligible signals in aortas of apoE+/+ mice with/without probe administration or atherosclerotic apoE-/- aortas without probe. Gelatinase inhibitor treatment abolished NIRF signals in apoE-/- mouse aortas ex vivo. Sites of gelatinase activity visualized by NIRF colocalized with macrophage accumulation, immunoreactive MMP-2 and MMP-9, and gelatinolytic activity detected by in situ zymography. Furthermore, fluorescence molecular tomography indicated in vivo that atherosclerotic aortas of apoE-/- mice produced NIRF signals for gelatinase action, whereas aortas of apoE+/+ mice injected with the probe or apoE-/- aortas with no probe exhibited negligible NIRF signals. CONCLUSIONS: These results suggest the feasibility of noninvasively imaging the enzymatic action of MMPs in vivo, an approach that may gauge inflammatory foci in atherosclerosis, assess cardiovascular risk, and evaluate the effects of therapeutic interventions.

Animals↗

From finite to infinite volumes: removal of boundaries in diffuse wave imaging.

In this Letter, we present a method that removes the contribution of the boundaries on the measurements from highly scattering media, transforming the signals captured from a bounded medium to measurements that would have been obtained if no boundary were present. This approach opens new possibilities in tomographic imaging in diffuse media as it eliminates the need for explicitly modeling boundaries and significantly simplifies reconstruction requirements.

Journal Article↗

Fluorescence molecular imaging.

There is a wealth of new fluorescent reporter technologies for tagging of many cellular and subcellular processes in vivo. This imposed contrast is now captured with an increasing number of available imaging methods that offer new ways to visualize and quantify fluorescent markers distributed in tissues. This is an evolving field of imaging sciences that has already achieved major advances but is also facing important challenges. It is nevertheless well poised to significantly impact the ways of biological research, drug discovery, and clinical practice in the years to come. Herein, the most pertinent technologies associated with in vivo noninvasive or minimally invasive fluorescence imaging of tissues are summarized. Focus is given to small-animal imaging. However, while a broad spectrum of fluorescence reporter technologies and imaging methods are outlined, as necessary for biomedical research, and clinical translation as well.

Animals↗

Normalized transillumination of fluorescent proteins in small animals.

We investigated the capacity of normalized transillumination methods in imaging fluorescent proteins in visible light, in phantoms and in live mice. We demonstrate significant imaging improvements over epi-illumination imaging, as commonly applied today. Of particular importance is the significant betterment in contrast achieved, due to minimization of the influence of autofluorescence and the enhancement of depth sensitivity compared to epi-illumination imaging. Although normalized transillumination does not match the accuracy and quantification capacity of tomographic imaging, it nevertheless attains key advantages over other planar optical imaging methods while preserving implementation simplicity. Due to the balance achieved between simplicity and accuracy, normalized transillumination approaches could serve as an important alternative molecular imaging method.

Animals↗

Volumetric tomography of fluorescent proteins through small animals in vivo.

Volumetric detection and accurate quantification of fluorescent proteins in entire animals would greatly enhance our ability to monitor biological processes in vivo. Here we present a quantitative tomographic technique for visualization of superficial and deep-seated (>2-3 mm) fluorescent protein activity in vivo. We demonstrate noninvasive imaging of lung tumor progression in a murine model, as well as imaging of gene delivery using a herpes virus vector. This technology can significantly improve imaging capacity over the current state of the art and should find wide in vivo imaging applications in drug discovery, immunology, and cancer research.

Animals↗

Use of gene expression profiling to direct in vivo molecular imaging of lung cancer.

Using gene expression profiling, we identified cathepsin cysteine proteases as highly up-regulated genes in a mouse model of human lung adenocarcinoma. Overexpression of cathepsin proteases in these lung tumors was confirmed by immunohistochemistry and Western blotting. Therefore, an optical probe activated by cathepsin proteases was selected to detect murine lung tumors in vivo as small as 1 mm in diameter and spatially separated. We generated 3D maps of the fluorescence signal and fused them with anatomical computed tomography images to show a close correlation between fluorescence signal and tumor burden. By serially imaging the same mouse, optical imaging was used to follow tumor progression. This study demonstrates the capability for molecular imaging of a primary lung tumor by using endogenous proteases expressed by a tumor. It also highlights the feasibility of using gene expression profiling to identify molecular targets for imaging lung cancer.

Adenocarcinoma, Bronchiolo-Alveolar↗

Effects of background fluorescence in fluorescence molecular tomography.

Recent advances in optical imaging systems and systemically administered fluorescent probes have significantly improved the ways by which we can visualize proteomics in vivo. A key component in the design of fluorescent probes is a favorable biodistribution, i.e., localization only in the targeted diseased tissue, in order to achieve high contrast and good detection characteristics. In practice, however, there is always some level of background fluorescence present that could result in distorted or obscured visualization and quantification of measured signals. In this study we observe the effects of background fluorescence in tomographic imaging. We demonstrate that increasing levels of background fluorescence result in artifacts when using a linear perturbation algorithm, along with a significant loss of image fidelity and quantification accuracy. To correct for effects of background fluorescence, we have applied cubic polynomial fits to bulk raw measurements obtained from spatially homogeneous and heterogeneous phantoms. We show that subtraction of the average fluorescence response from the raw data before reconstruction can improve image quality and quantification accuracy as shown in relatively homogeneous or heterogeneous phantoms. Subtraction methods thus appear to be a promising route for adaptively correcting nonspecific background fluorochrome distribution.

Algorithms↗

Tomographic fluorescence mapping of tumor targets.

Methods that allow robust imaging of specific molecular targets and biological processes in vivo should have widespread applications in biology and clinical medicine. Here we use a quantitative, three-dimensional fluorescence-mediated tomographic technique (FMT) that enables rapid measurements of fluorochrome-based affinity tags in live xenograft models. We validate the method by showing its sensitivity in quantitating tumor angiogenesis and therapeutic modulation using an anti-vascular endothelial growth factor antibody. Furthermore, we show the feasibility of simultaneous multichannel measurements of distinct biological phenomena such as receptor tyrosine kinase expression and angiogenesis. FMT measurements can be done serially, with short imaging times and within the same live animal. The described method should be valuable for rapidly profiling biological phenomena in vivo.

Animals↗

Quantitative point source photoacoustic inversion formulas for scattering and absorbing media.

We present here an expression for the photoacoustic contribution of an optical point source in a diffusive and absorbing medium. By using this measurement as a reference, we present a direct inversion formula that recovers the absorption map quantitatively, at the same time accounting for instrumental factors such as the source strength, the shape of the optical pulse, and the impulse response and finite size of the transducers. We further validate this expression through accurate numerical simulations showing that the absorption map is recovered quantitatively in the presence of a rotating geometry. We finally discuss how the presented solutions for point sources within the photoacoustic problem enable the use of concurrent fluorescence and ultrasound measurements as appropriate for a hybrid tomographic system. The proposed system could retrieve absorption information using photoacoustic measurements, and use these data to more accurately describe the fluorescence problem and improve reconstruction fidelity.

Absorption↗

Complete-angle projection diffuse optical tomography by use of early photons.

We present the first, to our knowledge, experimental images of complex-shaped phantoms embedded in diffuse media by use of optical tomography. Imaging is based on a complete-angle projection tomographic technique that utilizes transmitted early photons. Results are contrasted with measurements obtained at later gates as well as pseudocontinuous-wave data. The scanning system developed employs noncontact illumination and detection technologies that allow for high spatial sampling of transmitted photons. Combining this system with complete-angle illumination is found to be an important strategy toward improved imaging performance, resulting in a better-posed inversion problem. The appropriateness of reconstruction algorithms similar to those employed in x-ray computed tomography are showcased, and suggestions for model improvements are provided.

Algorithms↗

Looking and listening to light: the evolution of whole-body photonic imaging.

Optical imaging of live animals has grown into an important tool in biomedical research as advances in photonic technology and reporter strategies have led to widespread exploration of biological processes in vivo. Although much attention has been paid to microscopy, macroscopic imaging has allowed small-animal imaging with larger fields of view (from several millimeters to several centimeters depending on implementation). Photographic methods have been the mainstay for fluorescence and bioluminescence macroscopy in whole animals, but emphasis is shifting to photonic methods that use tomographic principles to noninvasively image optical contrast at depths of several millimeters to centimeters with high sensitivity and sub-millimeter to millimeter resolution. Recent theoretical and instrumentation advances allow the use of large data sets and multiple projections and offer practical systems for quantitative, three-dimensional whole-body images. For photonic imaging to fully realize its potential, however, further progress will be needed in refining optical inversion methods and data acquisition techniques.

Image Enhancement↗

Optical imaging of spontaneous breast tumors using protease sensing 'smart' optical probes.

OBJECTIVE: The objective of this study was to determine if spontaneous breast cancer lesions can be detected by fluorescence reflectance imaging (FRI) and fluorescence mediated tomography (FMT) using protease-sensing optical probes. MATERIALS AND METHODS: Transgenic (FVB/N-TgN (WapHRAS)69Lin Y)) mice, which spontaneously develop breast cancer, were injected intravenously with a cathepsin-sensing fluorescent imaging probe. FRI and FMT were performed 24 hours after probe injection and region of interest (ROI) analysis was performed. Magnetic resonance images were acquired for anatomic coregistration with the FMT data. Moreover, correlative immunohistochemistry and fluorescence microscopy were performed. RESULTS: All tumor nodules were clearly delineated by FRI showing an average signal intensity of 380 +/- 106 AU. Similarly, tumors were clearly detected by FMT imaging. Immunohistochemistry confirmed cathepsin-B expression of primary tumors and fluorescence microscopy revealed a strong Cy 5.5 deposition in the tissue. CONCLUSIONS: FRI and FMT using "smart" protease sensing probes permits detection of experimental spontaneous breast cancers. Because the expression levels of various proteases correlate with patient outcome, this technique may not only help to detect, but also to differentiate breast cancers noninvasively.

Animals↗

Fluorescent protein tomography scanner for small animal imaging.

Microscopy of fluorescent proteins has enabled unprecedented insights into visualizing gene expression in living systems. Imaging deeper into animals, however, has been limited due to the lack of accurate imaging methods for the visible. We present a novel system designed to perform tomographic imaging of fluorescent proteins through whole animals. The tomographic method employed a multiangle, multiprojection illumination scheme, while detection was achieved using a highly sensitive charge-coupled device camera with appropriate filters. Light propagation was modeled using a modified solution to the diffusion equation to account for the high absorption and high scattering of tissue at the visible wavelengths. We show that the technique can quantitatively detect fluorescence with sub millimeter spatial resolution both in phantoms and in tissues. We conclude that the method could be applied in tomographic imaging of fluorescent proteins for in vivo targeting of different diseases and abnormalities.

Animals↗

Image analysis for assessing molecular activity changes in time-dependent geometries.

In vivo fluorescence molecular imaging and tomography has facilitated monitoring of genomics and proteomics over time and on the same animal. A highly important issue, however, has been the robust registration of animals imaged at different time points to obtain accurate description of activity and location. This paper presents a method for aligning temporal data of small animals based on surface anatomical features and improving the accuracy of monitoring fluorophore distribution. The method can account for differences in the positioning and compression of small animals and can be extended to three-dimensional as well as to other imaging modalities.

Algorithms↗

Accuracy of fluorescent tomography in the presence of heterogeneities: study of the normalized Born ratio.

We studied the performance of three-dimensional fluorescence tomography of diffuse media in the presence of heterogeneities. Experimental measurements were acquired using an imaging system consisting of a parallel plate-imaging chamber and a lens coupled charge coupled device camera, which enables conventional planar imaging as well as fluorescence tomography. To simulate increasing levels of background heterogeneity, we employed phantoms made of a fluorescent tube surrounded by several absorbers in different combinations of absorption distribution. We also investigated the effect of low absorbing thin layers (such as membranes). We show that the normalized Born approach accurately retrieves the position and shape of the fluorochrome even at high background heterogeneity. We also demonstrate that the quantification is relatively insensitive to a varying degree of heterogeneity and background optical properties. Findings are further contrasted to images obtained with the standard Born expansion and with a normalized approach that divides the fluorescent field with excitation measurements through a homogeneous medium.

Algorithms↗

Quantitative analysis of chemotherapeutic effects in tumors using in vivo staining and correlative histology.

AIMS: To microscopically analyze the chemotherapeutic response of tumors using in vivo staining based on an annexinV-Cy5.5 probe and independently asses their apoptotic count using quantitative histological analysis. METHODS: Lewis Lung Carcinomas cells, that are sensitive (CS-LLC) and resistant (CR-LLC) to chemotherapy were implanted in nude mice and grown to tumours. Mice were treated with cyclophosphamide and injected with a Cy5.5-annexinV fluorescent probe. In vivo imaging was performed using Fluorescence Molecular Tomography. Subsequently tumours were excised and prepared for histology. The histological tumour sections were stained for apoptosis using a terminal deoxynucleotidyl transferase-mediated nick end labeling (TUNEL) assay. A minimum of ten tissue sections were analyzed per tumour for apoptosis quantification by TUNEL staining and corresponding Cy5.5 distribution. RESULTS: We detected higher levels of apoptosis and corresponding higher levels of Cy5.5 fluorescence in the CS-LLC vs. the CR-LLC tumours. The cell count rate on CS-LLC sections over CR-LLC was found to be approximately 2 :1 where the corresponding area observed on Cy5.5 distribution measurements revealed a approximately 1.7 :1 ratio of CS-LLC over CR-LLC. These observations are consistent with the higher apoptotic index expected from the CS-LLC cell line. CONCLUSIONS: Quantitative analysis of histological slices revealed higher fluorescence and higher apoptotic count in the CS-LLC tumour images compared to the CR-LLC tumour images. These observations demonstrate that the annexinV-Cy5.5 probe sensed the chemotherapeutic effect of cyclophospamide and further confirmed in vivo FMT measurements.

Animals↗

Experimental determination of photon propagation in highly absorbing and scattering media.

Optical imaging and tomography in tissues can facilitate the quantitative study of several important chromophores and fluorophores. Several theoretical models have been validated for diffuse photon propagation in highly scattering and low-absorbing media that describe the optical appearance of tissues in the near-infrared (NIR) region. However, these models are not generally applicable to quantitative optical investigations in the visible because of the significantly higher tissue absorption in this spectral region compared with that in the NIR. We performed photon measurements through highly scattering and absorbing media for ratios of the absorption coefficient to the reduced scattering coefficient ranging approximately from zero to one. We examined experimentally the performance of the absorption-dependent diffusion coefficient defined by Aronson and Corngold [J. Opt. Soc. Am. A 16, 1066 (1999)] for quantitative estimations of photon propagation in the low- and high-absorption regimes. Through steady-state measurements we verified that the transmitted intensity is well described by the diffusion equation by considering a modified diffusion coefficient with a nonlinear dependence on the absorption. This study confirms that simple analytical solutions based on the diffusion approximation are suitable even for high-absorption regimes and shows that diffusion-approximation-based models are valid for quantitative measurements and tomographic imaging of tissues in the visible.

Animals↗