Non-contact tonometry over soft contact lenses: effect of contact lens power on the measurement of intra-ocular pressure.
PURPOSE: To perform non-contact applanation tonometry over soft contact lenses and to examine any differences between the measured IOPs with and without the contact lens. METHOD: IOP was measured at one eye of eight subjects and re-evaluated through a soft lens in situ over the cornea. Seventy percent (ES70, Lunelle) water contact lenses or powers ranging from -15.00 D to +13.00 D were used. RESULTS: Mean IOP without lenses was 15.2 mmHg (S.D. = +/-4.2) and this did not change significantly for lenses ranging from -15.00 D to +3.00 D (P > 0.01) but did change for lens powers +6.00 D and above (P<0.01). The least squares line describing the change in measured IOP (deltaIOP) as a function of lens power (x) has the following characteristics, deltaIOP = 0.022 + 0.591x + 0.11x2 + 0.005x3 (r = 0.985, P = 0.0001). CONCLUSION: Non-contact tonometry can be performed with sufficient accuracy over a soft lens on condition: (a) lens centre thickness is no more 0.30 mm and (b) power is not greater than +3D.