PubMed HealthSearch

Biomedical subjects

W W Peppler

Publications and source records attributed to W W Peppler.

At least 19 recordsLinked to original sources

Imaging characteristics of x-ray capillary optics in digital mammography.

Computed radiography (CR) has shown promise in digital mammographic screening due to its good low spatial frequency MTF and its relatively wide exposure latitude. The CR image format has not gained acceptance clinically because of reduced high spatial frequency resolution as compared to film-screen images. X-ray capillary optics, aligned between the breast and CR phosphor imaging plate, will capture primary x-ray photons almost exclusively. Due to the very small angle of acceptance, scattered photons angled more than about 1.6 x 10(-3) radians from primary trajectory will not be accepted at the capillary optic entrance. The virtual elimination of detected scatter means almost 100% of the possible primary contrast should be visible in the image. In addition, the image can be magnified without focal spot blurring. Effective resolution of CR images can be increased by a factor equal to that magnification. Clinical implementation of future capillary optics are expected to be either in the form of a large, stationary, post-patient optic that accepts primary from the entire breast or a fan-shaped optic that is scanned across the breast. Measurements of a test capillary optic showed a reduction of scatter fraction to 0.018. Images of a lucite contrast detail phantom revealed a corresponding increase in image contrast when compared to anti-scatter grid and no grid methods. Spectral transmission measurements using a high-purity germanium detector showed good primary transmission (45%-50%) in the mammographic energy range. The MTF measurements of both stationary and scanned capillary optics showed improvement at the 5% MTF level to 8.4 mm-1 for scanned optics and 9.2 mm-1 for stationary optics representing a 68% and 84% respective increase over the CR MTF without magnification or capillary optics.

Capillary Action

Technical exhibits.

Explore the source record for details and available documents.

Exhibitions as Topic

Measurements of capillary x-ray optics with potential for use in mammographic imaging.

Capillary optic arrays are bundles of hollow glass capillaries which guide x rays in a manner similar to the way fiber optics guide light. Focused postpatient capillary optic arrays have the potential to significantly improve both contrast and resolution of mammographic images compared to conventional antiscatter grids. Contrast can be improved by the nearly total scatter rejection of the optic. Effective resolution can be improved by geometric magnification without increased focal spot blurring. The best results were found for borosilicate glasses, with transmissions in excess of 60% for 22-cm-long fibers. To evaluate the scatter rejection properties, the transmission of off-axis radiation was measured. Transmission drops to < 1% at an angular displacement of 2.7 mrad. Transmission of a bulk capillary array dropped to near zero if the source was at an angle of 2.5 mrad. This implies excellent scatter rejection capabilities. To evaluate whether unchanneled photons might still reach the detector, absorption measurements were also performed on fibers and arrays. Absorption was found to be adequate for scatter rejection. All of the data agreed well with numerical simulations. Performance calculations for two potential optics geometries gave promising results.

Biophysical Phenomena

A regional convolution kernel algorithm for scatter correction in dual-energy images: comparison to single-kernel algorithms.

Single kernel scatter correction algorithms are based on the model that the scatter field can be predicted by convolution of the primary intensity (Iprim) with a spatially invariant scatter point-spread function (PSF). Practical limitations (Iprim unknown) suggest the substitution of the total detected intensity (Idet) for Iprim as the source image in the convolution. In regions of high scatter fraction (SF), Idet is a poor approximation of Iprim, thereby causing an overestimation of scatter originating in the region. This contributes to errors in estimating detected scatter in the mediastinum and neighboring regions. A technique using a regionally variable point-spread function that significantly reduces RMS error in estimation of the primary image as compared to the single PSF method is investigated. The regionally variable convolution method employs a larger PSF in the mediastinum and a smaller PSF in the lungs to reduce the error in estimating the scatter throughout the image. The method to allow for patient differences has also been expanded and various implementations of these methods have been compared. Results show that the dual-kernel algorithm is always more effective than an equivalent single-kernel algorithm. The dual-kernel algorithm using a predicted scatter fraction curve gives an overall RMS error in the primary of as low as 20.8% which is equivalent to 8.7% RMS error in the scatter. The dual-kernel method using a predicted scatter fraction curve approaches the accuracy of the single-kernel method using patient specific scatter measurements. Because using individual scatter measurements is a less desirable method for clinical use, we feel that the dual-kernel algorithm which uses two regions specific convolution kernels and a variable scatter fraction curve is the preferable method.

Algorithms

Technical exhibits.

Explore the source record for details and available documents.

Diagnostic Imaging

Conventional chest radiography vs dual-energy computed radiography in the detection and characterization of pulmonary nodules.

OBJECTIVE: We evaluated a single-exposure, phosphor-plate, dual-energy imaging device that produces, in addition to conventional chest radiographs, both tissue- and bone-selective images. Our purpose was to determine whether dual-energy radiography was more accurate than routine chest radiography for detection and characterization of pulmonary nodules. SUBJECTS AND METHODS: Two hundred patients undergoing chest CT were asked to volunteer to have dual-energy and conventional chest radiographs obtained immediately before or after their CT scan. Radiographs from a subset of 50 of these patients with 116 CT-detected nodules and 10 patients with normal findings on CT scans of the chest were presented to the observers for the nodule detection study. Similarly, radiographs from a subset of 29 patients with 20 calcified and 20 uncalcified nodules were presented to five observers to determine nodule calcification. Dual-energy images were produced by filtering the X-ray tube output with a gadolinium sheet while using a multiple phosphor plate receptor. A dual-energy triad of images consisting of a conventional image, a tissue-selective image, and a bone-selective image were produced. The conventional chest radiographs and dual-energy image sets were presented to observers in random order. Data from a free response receiver operating curve and a receiver operating curve were generated for nodule detection and characterization, respectively. RESULTS: By using the dual-energy images, all five observers improved their ability to diagnose pulmonary nodules (p = .0005) and to characterize nodules as calcified (p = .005). CONCLUSION: By eliminating rib shadows with tissue-selective images and enhancing calcified structures with bone-selective images, dual-energy chest radiography improved the ability of all observers, regardless of expertise, to detect and characterize pulmonary nodules.

False Positive Reactions

Densitometric assessment of regional left ventricular systolic function during graded ischemia in the dog by use of dual-energy digital subtraction ventriculography.

Densitometric analysis of images obtained by digital subtraction angiography (DSA) allows for more reproducible and less operator-dependent quantitation of ventricular function. Conventional DSA uses temporal subtraction but is limited by misregistration artifacts. Dual-energy digital subtraction angiography (DE-DSA) is immune to such misregistration artifacts. The ability of DE-DSA to quantitate changes in regional ventricular volume resulting from ischemia was tested. Densitometric analysis of both phase-matched and ejection fraction DE-DSA images was used to quantitate regional left ventricular systolic function during four levels of ischemia ranging from mild to severe in open-chest dogs (n = 10). DE-DSA left ventriculograms were obtained by means of central venous injections of iodinated contrast medium. Ischemia was graded according to percentage of systolic wall thickening as measured by sonomicrometry. Phase-matched end-systolic images were obtained at each of four levels of ischemia by subtracting an end-systolic control image from each end-systolic ischemic image. Ejection fraction images were obtained at the control level and at each level of ischemia by subtracting an end-systolic image from an end-diastolic image of the same cardiac cycle. The resulting wall motion difference signals represent the changes in regional ventricular volumes and were quantitated by densitometry. Densitometry was able to detect the effect of all levels of ischemia on regional function, even the mildest. Densitometric analysis of both phase-matched and ejection fraction DE-DSA images provides a sensitive technique for detecting and quantitating the changes in regional left ventricular systolic volume that occur with ischemia.

Angiography, Digital Subtraction

Geometric quantitative coronary arteriography. A comparison of unsubtracted and dual energy-subtracted images.

The application of dual energy (DE) subtraction techniques to quantitative coronary arteriography (QCA) has the advantage of removing the tissue signal surrounding the vessel profile. We have compared the performance of two geometric QCA algorithms on DE-subtracted and -unsubtracted images to determine, for each, if DE subtraction is advantageous. The two algorithms under study were an edge detection algorithm and a Fourier analysis-based algorithm. For each algorithm, linear regression analysis was performed of measured cross-sectional area (CSA) versus actual CSA of coronary vessel phantoms. The edge detection algorithm was found to have improved precision (P less than .05) when applied to the DE-subtracted images. The Fourier analysis algorithm, however, was not effected by the DE subtraction. Among the unsubtracted image results, the Fourier measurements were more accurate (P less than .05) than the edge detection measurements. We conclude that the benefits to edge detection QCA of DE tissue subtraction outweigh the disadvantages of increased image noise and possible misregistration artifacts. However, the Fourier algorithm is relatively insensitive to tissue signal variations.

Algorithms

Quantitative dual-energy coronary arteriography.

Subtraction techniques for digital cardiac imaging have been hampered by misregistration artifacts. The use of dual-energy imaging is being evaluated as a means for reducing these artifacts. Results reported previously indicate that the dual-energy technique may be useful for applications such as exercise ventriculography and general quantification tasks. The purpose of the current study is to investigate the use of dual-energy subtraction imaging for quantitative coronary arteriography. In vivo coronary vessel phantoms (0.2 to 7 mm2 in cross-sectional area) were used to study the potential advantages of tissue suppressed energy subtracted images over unsubtracted images for quantification of absolute vessel cross-sectional area when cardiac motion is present. Estimates of lumen cross-sectional area (N = 20) were determined using videodensitometric analysis of selected energy subtracted and unsubtracted images. Linear regression analysis of measured and actual cross-sectional area showed energy subtracted image data (slope = 1.06, intercept = 0.48 mm2, r = 0.99) to have improved accuracy (P less than .05) and precision (P less than .05) over unsubtracted image data (slope = 1.24, intercept = 1.07 mm2, r = 0.95).

Angiography, Digital Subtraction

K-edge digital subtraction arthrography of the painful hip prosthesis: a feasibility study.

K-edge energy subtraction radiography is a method for detecting the presence of iodinated contrast material by subtracting two digital radiographs produced by X-ray beams with energies above and below the iodine K edge. We performed a feasibility study on the application of K-edge energy digital subtraction arthrography (KEDSA) to painful hip prostheses. During arthrography, loosening of the prosthesis is implied if contrast material is seen dissecting around the prosthesis, an often difficult detection task because of adjacent prosthesis metal or cement. In conventional arthrography a preliminary mask image is thus used from which films obtained after injection of iodinated contrast material are subtracted. Movement by the patient during this process may preclude subsequent subtraction. With KEDSA, since multiple image pairs may be obtained after the injection of contrast material, the problem of patient motion is virtually eliminated. A conventional X-ray tube operating between 55 and 65 kVp was alternately filtered by iodine and cerium filters to produce the KEDSA images. The apparatus was capable of producing a subtracted image within 3 sec. The technique was applied to phantoms and to six patients immediately after hip arthrography that had been positive for prosthesis loosening. Although of lower spatial resolution, the KEDSA images were, in all cases, positive for loosening in a pattern consistent with the conventional arthrographic images. KEDSA was shown to be successful in detecting extraarticular contrast material. During a single study, subtraction in various imaging planes as well as postexercise subtraction imaging can be accomplished-techniques not heretofore possible in routine subtraction arthrography.

Aged

Digital subtraction angiographic imaging of coronary flow reserve.

Recent studies have demonstrated that subjective assessment of the severity of coronary artery stenoses results in poor interobserver concordance and poor correlation with physiologic significance as determined by Doppler measurements of coronary flow reserve. Use of the coronary flow reserve as an integrated measure of the effect of stenosis geometry has been emphasized within the context of quantitative cinemetric analysis. The comparison of two parametric digital subtraction angiographic flow images obtained before and after hyperemic intervention has led to calculation of flow reserve values that correlate well with electromagnetic flowmeter data in dogs. By means of a similar model relating blood flow and image variables, single flow ratio images have been formed. These parametric images provide a two-dimensional display of the ratio of hyperemic flow to baseline flow. Linear temporal interpolation of data from a sequence of cardiac phase-matched subtraction images is used to improve the resolution of the displayed flow ratios. Summation of flow variables measured within the perfusion bed was used to calculate a value for the overall coronary flow reserve and to characterize the significance of isolated lesions in an open-chest canine preparation. A linear regression calculation relating parametric image flow ratio values to electromagnetic flowmeter measurements resulted in a linear fit of y = .96x - 0.19 with a correlation coefficient of .90. The direct visual representation of flow ratio distribution provided by the parametric imaging method may aid in the interpretation of multiple complex lesions as well as of single lesions.

Angiography

Digital beam attenuator technique for compensated chest radiography.

The feasibility of producing patient-specific beam attenuators for chest radiography has been investigated using an anthropomorphic phantom and a human volunteer. A low-dose test exposure is digitized, processed, and used to print a small cerium filter, which is placed in the x-ray beam near the collimator. The final radiograph is recorded on film. The technique results in relatively uniform film exposure, so that structures in all regions of the chest are simultaneously displayed with optimal film contrast. The equalized exposure improves image quality in the normally underpenetrated regions and reduces the role of cross-scatter from the lungs. The image is analogous to optical or computer-processed unsharp masking techniques, but the processing is accomplished in the x-ray beam and results in an improved exposure distribution, giving advantages that cannot be achieved with image processing techniques alone.

Humans

Total body and regional bone mineral by dual-photon absorptiometry in metabolic bone disease.

Dual-photon absorptiometry (153Gd) was used to measure bone mineral of the total body and major anatomical areas. Patients with osteoporosis (male = 11, female = 18) and with renal osteodystrophy (n = 17) were significantly below (20%) normal females (n = 72) and males (n = 13) at most sites. In the osteoporotic patients, but not the renal patients, there was preferential osteopenia of the spine. Bone loss in all anatomical areas became evident after the menopause with an annual loss rate of about 0.7%.

Absorption

Total body composition by dual-photon (153Gd) absorptiometry.

The lean-fat composition (%FATR) of soft tissue and the mineral mass of the skeleton were determined in vivo using dual-photon (153Gd) absorptiometry (dose under 2 mrem). A rectilinear raster scan was made over the entire body in 18 subjects (14 female, 4 male). Single-photon absorptiometry (125I) measured bone mineral content on the radius. Percentage fat (%FATD) was determined in the same subjects using body density (from underwater weighing with correction for residual lung volume). Lean body mass (LBM) was determined using both %FATR and %FATD. Percentage fat from absorptiometry and from underwater density were correlated (r = 0.87). The deviation of %FATD from %FATR was due to the amount of skeletal mineral as a percentage of the LBM (r = 0.90). Therefore, skeletal variability, even in normal subjects, where mineral ranges only from 4 to 8% of the LBM, essentially precludes use of body density as a composition indicator unless skeletal mass is measured. Anthropometry (fatfolds and weight) predicted %FATR and LBM at least as well as did underwater density. The predictive error of %FATR from fatfolds was 5% while the predictive error in predicting LBM from anthropometry was 2 to 3 kg (3%).

Adipose Tissue

Does bone measurement on the radius indicate skeletal status? Concise communication.

Single-photon (I-125) absorptiometry was used to measure bone mineral content (BMC) of the distal third of the radius, and dual-photon absorptiometry (Gd-153) was used to measure total-body bone mineral (TBBM), as well as the BMC of major skeletal regions. Measurements were done in normal females, normal males, osteoporotic females, osteoporotic males, and renal patients. The BMC of the radius predicted TBBM well in normal subjects, but was less satisfactory in the patient groups. The spinal BMC was predicted with even lower accuracy from radius measurement. The error in predicting areal density (bone mass per unit projected skeletal area) of the lumbar and thoracic spine from the radius BMC divided by its width was smaller, but the regressions differed significantly among normals, osteoporotics, and renal patients. There was a preferential spinal osteopenia in the osteoporotic group and in about half of the renal patients. Bone measurements on the radius can indicate overall skeletal status in normal subjects and to a lesser degree in patients, but these radius measurements are inaccurate, even on the average, as an indicator of spinal state.

Adult

Videodensitometric quantitation of mean blood flow.

Standard angiography demonstrates the anatomy of arterial occlusive disease but does not define its physiological significance. However, measurement of flow in a compromised vessel at rest and following peripheral dilatation provides important physiological information. Using digital subtraction angiography, femoral arterial flows determined by the cross-correlation transit time technique were compared to measurements by electromagnetic flowmeter. Thirty-five femoral arterial flow measurements were obtained in nine dogs instrumented with an electromagnetic flow probe and balloon occluder. Renografin 76 (7 cc) was power-injected at 14 cc/sec into the distal abdominal aorta. Angiographic flow measurements correlated well with electromagnetic flowmeter measurements (r = 0.94, standard deviation of the difference (SDD) = 15 ml/min). Intravenous studies provided somewhat poorer correlation due to difficulties in defining dimensions (r = 0.72, SDD = 36). Paired contrast injections (2 injections in succession) in 11 studies increased flow from an average of 80 to 250 ml/min (a 210 +/- 100% increase), providing an estimate of a vessel's capacity to provide increased flow during peripheral dilatation. Thus, reliable angiographic flow determinations may be obtained by arterial and intravenous contrast injections, adding physiological information to anatomical definition.

Absorptiometry, Photon

Work in progress: hybrid temporal-energy subtraction in digital fluoroscopy.

Initial clinical results using a digital fluoroscopic implementation of the combined time-energy ("hybrid") subtraction technique are described, with emphasis on carotid and renal imaging. Where patient motion artifacts are due to soft-tissue motion alone, hybrid subtraction can remove them. Due to the need for a finite separation time between high- and low-energy pairs, however, the present implementation of the hybrid technique is not completely immune to soft-tissue motion. The intrinsic signal-to-noise ratio of hybrid imaging is less than that of conventional temporal subtraction. However, since the low-energy temporal subtraction images are included in the hybrid data set, the diagnostic quality of the examination is not compromised.

Analog-Digital Conversion