Common-path interferometer for frequency-domain optical coherence tomography.

A Michelson-type spectral interferometer that uses a common beam path for the reference and the sample arms is described. This optical arrangement is more compact and stable than the more commonly used dual-arm interferometer and is well suited for frequency-domain optical coherence tomography of biological samples. With a 16-bit CCD camera, the instrument has sufficient dynamic range and resolution for imaging to depths of 2 mm in scattering biological materials. Images obtained with this spectral interferometer are presented, including cross-sectional images in a Xenopus laevis tadpole.

Abdomen↗

Differential spectral interferometry: an imaging technique for biomedical applications.

Differential spectral interferometry (DSI), a novel method of biomedical imaging that combines the high dynamic range of optical coherence tomography (OCT) with inherently parallel low-bandwidth image acquisition of spectral interferometry (SI), is described. DSI efficiently removes the deleterious dc background inherent in SI measurements while maintaining the parallel nature of SI. We demonstrate DSI on both synthetic and biological samples. Because DSI preserves the low-bandwidth, parallel nature of SI, it is competitive with OCT for biomedical applications in terms of image quality and acquisition rate.

Anatomy, Cross-Sectional↗