PubMed Health⌕ Search

Biomedical subjects

Wolfgang Stockmann

Publications and source records attributed to Wolfgang Stockmann.

2 recordsLinked to original sources

Multicentre performance evaluation of the E170 module for modular analytics.

The E170 module was evaluated at 13 sites in an international multicentre study. The objective of the study was to assess the analytical performance of 49 analytes, and to collect feedback on the system's reliability and practicability. The typical, within-run coefficients of variation (CVs) for most of the quantitative assays ranged between 1 and 2% while a range of 2-4% was achieved with the infectious disease methods. Total precision CVs were found to be within the manufacturer's expected performance ranges, demonstrating good concordance of the system's measuring channels and a high reproducibility during the 2-4-week trial period. The functional sensitivity of 11 selected assays met the clinical requirements (e.g., thyreotroponin (TSH) 0.008 mU/l, troponin T 0.02 microg/l, total prostate-specific antigen (PSA) 0.03 microg/l). The E170 showed no drift during an 8-hour period and no relevant reagent carryover. Accuracy was confirmed by ring trial experiments and method comparisons vs. Elecsys 2010. The reliability and practicability of the system's hardware and software met with, or even exceeded, the evaluator's requirements. Workflow studies showed that E170 can cover the combined workload of various routine analysers in a variety of laboratory environment. Throughput and sample processing time requirements were achieved while personnel 'hands-on-time' could be reduced.

Calibration↗

Evaluation of Cobas Integra 800 under simulated routine conditions in six laboratories.

The new selective access analyser Cobas Integra 800 from Roche Diagnostics was evaluated in an international multicentre study at six sites. Routine simulation experiments showed good performance and full functionality of the instrument and provocation of anomalous situations generated no problems. The new features on Cobas Integra 800, namely clot detection and dispensing control, worked according to specifications. The imprecision of Cobas Integra 800 fulfilled the proposed quality specifications regarding imprecision of analytical systems for clinical chemistry with few exceptions. Claims for linearity, drift, and carry-over were all within the defined specifications, except urea linearity. Interference exists in some cases, as could be expected due to the chemistries applied. Accuracy met the proposed quality specifications, except in some special cases. Method comparisons with Cobas Integra 700 showed good agreement; comparisons with other analysis systems yielded in several cases explicable deviations. Practicability of Cobas Integra 800 met or exceeded the requirements for more than 95% of all attributes rated. The strong points of the new analysis system were reagent handling, long stability of calibration curves, high number of tests on board, compatibility of the sample carrier to other Roche systems, and the sample integrity check for more reliable analytical results. The improvement of the workflow offered by the 5-position rack and STAT handling like on Cobas Integra 800 makes the instrument attractive for further consolidation in the medium-sized laboratory, for dedicated use of special analytes, and/or as back-up in the large routine laboratory.

Blood Proteins↗