Precision wavelength measurement of the 14.4 keV Mössbauer photon.

The wavelength of the 14.4 keV Mössbauer photon has been determined by using undulator radiation and diffraction of FZ silicon crystals. For the wavelength determination a goniometer equipped with a laser rotary encoder and a sine-bar angle optical interferometer was developed; a temperature-monitoring system and X-ray optics related to the experiment were also developed. The mean wavelength was 0.08602557 nm with an uncertainty of 0.6 p.p.m., derived from 16 measurements of three pieces of FZ Si(840) crystals.

Journal Article↗

Precision goniometer equipped with a 22-bit absolute rotary encoder.

The calibration of a compact precision goniometer equipped with a 22-bit absolute rotary encoder is presented. The goniometer is a modified Huber 410 goniometer: the diffraction angles can be coarsely generated by a stepping-motor-driven worm gear and precisely interpolated by a piezoactuator-driven tangent arm. The angular accuracy of the precision rotary stage was evaluated with an autocollimator. It was shown that the deviation from circularity of the rolling bearing utilized in the precision rotary stage restricts the angular positioning accuracy of the goniometer, and results in an angular accuracy ten times larger than the angular resolution of 0.01 arcsec. The 22-bit encoder was calibrated by an incremental rotary encoder. It became evident that the accuracy of the absolute encoder is approximately 18 bit due to systematic errors.