PubMed Health⌕ Search

Biomedical subjects

Zhouping Wei

Publications and source records attributed to Zhouping Wei.

6 recordsLinked to original sources

Needle and seed segmentation in intra-operative 3D ultrasound-guided prostate brachytherapy.

In order to guide the needle to the correct location in 3D US-guided brachytherapy, the needle is continuously tracked as it is being inserted. A pre-scan before the needle insertion and a post-scan after the needle insertion are subtracted to obtain a difference image containing the needle. The image is projected along two orthogonal directions approximately perpendicular to the needle, and the 3D needle is reconstructed from the segmented needles in the two projected images. The seeds implanted with the needle are located in the cropped volume along the needle. Thus, the seeds are segmented using a tri-bar model and 3D line segment patterns. Finally, the positions of the seeds are determined using a peak detection technique. Experiments with agar and turkey/chicken phantoms as well as patient data demonstrated that our needle segmentation technique could segment the needle in near real-time with an accuracy of 0.6 mm in position and 1.0 degrees in orientation. The true-positive rate for seed segmentation is 100% for the agar phantom and 93% for the chicken phantom. The average distance to manual seed segmentation was 1.0mm for the agar phantom and 1.7 mm for the chicken phantom.

Algorithms↗

Automated localization of implanted seeds in 3D TRUS images used for prostate brachytherapy.

An algorithm has been developed in this paper to localize implanted radioactive seeds in 3D ultrasound images for a dynamic intraoperative brachytherapy procedure. Segmentation of the seeds is difficult, due to their small size in relatively low quality of transrectal ultrasound (TRUS) images. In this paper, intraoperative seed segmentation in 3D TRUS images is achieved by performing a subtraction of the image before the needle has been inserted, and the image after the seeds have been implanted. The seeds are searched in a "local" space determined by the needle position and orientation information, which are obtained from a needle segmentation algorithm. To test this approach, 3D TRUS images of the agar and chicken tissue phantoms were obtained. Within these phantoms, dummy seeds were implanted. The seed locations determined by the seed segmentation algorithm were compared with those obtained from a volumetric cone-beam flat-panel micro-CT scanner and human observers. Evaluation of the algorithm showed that the rms error in determining the seed locations using the seed segmentation algorithm was 0.98 mm in agar phantoms and 1.02 mm in chicken phantoms.

Algorithms↗

3D TRUS guided robot assisted prostate brachytherapy.

This paper describes a system for dynamic intraoperative prostate brachytherapy using 3D ultrasound guidance with robot assistance. The system consists of 3D transrectal ultrasound (TRUS) imaging, a robot and software for prostate segmentation, 3D dose planning, oblique needle segmentation and tracking, seed segmentation, and dynamic re-planning and verification. The needle targeting accuracy of the system was 0.79 mm +/- 0.32 mm in a phantom study.

Algorithms↗

Brachytherapy needle deflection evaluation and correction.

In prostate brachytherapy, an 18-gauge needle is used to implant radioactive seeds. This thin needle can be deflected from the preplanned trajectory in the prostate, potentially resulting in a suboptimum dose pattern and at times requiring repeated needle insertion to achieve optimal dosimetry. In this paper, we report on the evaluation of brachytherapy needle deflection and bending in test phantoms and two approaches to overcome the problem. First we tested the relationship between needle deflection and insertion depth as well as whether needle bending occurred. Targeting accuracy was tested by inserting a brachytherapy needle to target 16 points in chicken tissue phantoms. By implanting dummy seeds into chicken tissue phantoms under 3D ultrasound guidance, the overall accuracy of seed implantation was determined. We evaluated methods to overcome brachytherapy needle deflection with three different insertion methods: constant orientation, constant rotation, and orientation reversal at half of the insertion depth. Our results showed that needle deflection is linear with needle insertion depth, and that no noticeable bending occurs with needle insertion into the tissue and agar phantoms. A 3D principal component analysis was performed to obtain the population distribution of needle tip and seed position relative to the target positions. Our results showed that with the constant orientation insertion method, the mean needle targeting error was 2.8 mm and the mean seed implantation error was 2.9 mm. Using the constant rotation and orientation reversal at half insertion depth methods, the deflection error was reduced. The mean needle targeting errors were 0.8 and 1.2 mm for the constant rotation and orientation reversal methods, respectively, and the seed implantation errors were 0.9 and 1.5 mm for constant rotation insertion and orientation reversal methods, respectively.

Agar↗

Oblique needle segmentation and tracking for 3D TRUS guided prostate brachytherapy.

An algorithm was developed in order to segment and track brachytherapy needles inserted along oblique trajectories. Three-dimensional (3D) transrectal ultrasound (TRUS) images of the rigid rod simulating the needle inserted into the tissue-mimicking agar and chicken breast phantoms were obtained to test the accuracy of the algorithm under ideal conditions. Because the robot possesses high positioning and angulation accuracies, we used the robot as a "gold standard," and compared the results of algorithm segmentation to the values measured by the robot. Our testing results showed that the accuracy of the needle segmentation algorithm depends on the needle insertion distance into the 3D TRUS image and the angulations with respect to the TRUS transducer, e.g., at a 10 degrees insertion anglulation in agar phantoms, the error of the algorithm in determining the needle tip position was less than 1 mm when the insertion distance was greater than 15 mm. Near real-time needle tracking was achieved by scanning a small volume containing the needle. Our tests also showed that, the segmentation time was less than 60 ms, and the scanning time was less than 1.2 s, when the insertion distance into the 3D TRUS image was less than 55 mm. In our needle tracking tests in chicken breast phantoms, the errors in determining the needle orientation were less than 2 degrees in robot yaw and 0.7 degrees in robot pitch orientations, for up to 20 degrees needle insertion angles with the TRUS transducer in the horizontal plane when the needle insertion distance was greater than 15 mm.

Agar↗

Robot-assisted 3D-TRUS guided prostate brachytherapy: system integration and validation.

Current transperineal prostate brachytherapy uses transrectal ultrasound (TRUS) guidance and a template at a fixed position to guide needles along parallel trajectories. However, pubic arch interference (PAI) with the implant path obstructs part of the prostate from being targeted by the brachytherapy needles along parallel trajectories. To solve the PAI problem, some investigators have explored other insertion trajectories than parallel, i.e., oblique. However, parallel trajectory constraints in current brachytherapy procedure do not allow oblique insertion. In this paper, we describe a robot-assisted, three-dimensional (3D) TRUS guided approach to solve this problem. Our prototype consists of a commercial robot, and a 3D TRUS imaging system including an ultrasound machine, image acquisition apparatus and 3D TRUS image reconstruction, and display software. In our approach, we use the robot as a movable needle guide, i.e., the robot positions the needle before insertion, but the physician inserts the needle into the patient's prostate. In a later phase of our work, we will include robot insertion. By unifying the robot, ultrasound transducer, and the 3D TRUS image coordinate systems, the position of the template hole can be accurately related to 3D TRUS image coordinate system, allowing accurate and consistent insertion of the needle via the template hole into the targeted position in the prostate. The unification of the various coordinate systems includes two steps, i.e., 3D image calibration and robot calibration. Our testing of the system showed that the needle placement accuracy of the robot system at the "patient's" skin position was 0.15 mm+/-0.06 mm, and the mean needle angulation error was 0.07 degrees. The fiducial localization error (FLE) in localizing the intersections of the nylon strings for image calibration was 0.13 mm, and the FLE in localizing the divots for robot calibration was 0.37 mm. The fiducial registration error for image calibration was 0.12 mm and 0.52 mm for robot calibration. The target registration error for image calibration was 0.23 mm, and 0.68 mm for robot calibration. Evaluation of the complete system showed that needles can be used to target positions in agar phantoms with a mean error of 0.79 mm+/-0.32 mm.

Automation↗