Accuracy of numerically produced compensators.

Abstract

A feasibility study is performed to assess the utility of a computer numerically controlled (CNC) mill to produce compensating filters for conventional clinical use and for the delivery of intensity-modulated beams. A computer aided machining (CAM) software is used to assist in the design and construction of such filters. Geometric measurements of stepped and wedged surfaces are made to examine the accuracy of surface milling. Molds are milled and filled with molten alloy to produce filters, and both the molds and filters are examined for consistency and accuracy. Results show that the deviation of the filter surfaces from design does not exceed 1.5%. The effective attenuation coefficient is measured for CadFree, a cadmium-free alloy, in a 6 MV photon beam. The effective attenuation coefficients at the depth of maximum dose (1.5 cm) and at 10 cm in solid water phantom are found to be 0.546 cm-1 and 0.522 cm-1, respectively. Further attenuation measurements are made with Cerrobend to assess the variations of the effective attenuation coefficient with field size and source-surface distance. The ability of the CNC mill to accurately produce surfaces is verified with dose profile measurements in a 6 MV photon beam. The test phantom is composed of a 10 degrees polystyrene wedge and a 30 degrees polystyrene wedge, presenting both a sharp discontinuity and sloped surfaces. Dose profiles, measured at the depth of compensation (10 cm) beneath the test phantom and beneath a flat phantom, are compared to those produced by a commercial treatment planning system. Agreement between measured and predicted profiles is within 2%, indicating the viability of the system for filter production.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

H Thompson, M D Evans, B G Fallone. 1999. Accuracy of numerically produced compensators.. https://doi.org/10.1016/s0958-3947(98)00049-1

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

A feasibility study on the use of visible and short wavelengths in the near-infrared region for the non-destructive measurement of wine composition.

The aim of this study was to explore the capability of spectroscopy in the visible (Vis) and short wavelength near-infrared (NIR) regions for the non-destructive measurement of wine composition in intact bottles. In this study we analysed a wide range of commercial wines obtained in Australia in different types of bottles (e.g. colours, diameters and heights), including different wine styles and varieties. Wine bottles were scanned in the Vis-NIR region (600-1,100 nm) in a monochromator instrument in transflectance mode. Principal component analysis (PCA) and partial least-squares (PLS) regression were used to interpret the spectra and develop calibrations for wine composition. Due to the relatively small number of samples available full cross-validation (leave-one-out) was used as validation. The coefficient of correlation in calibration [Formula: see text] and the standard error of cross-validation (SECV) were 0.67 (SECV: 0.48%), 0.83 (SECV: 4.01 mg L-1), 0.70 (SECV: 28.6 mg L-1) and 0.50 (SECV: 0.15) for alcohol content, total SO2, free SO2 and pH, respectively, in the set of wine samples analysed. These preliminary results showed that the assessment of wine composition by Vis and short wavelengths in the NIR is possible for either qualitative analysis (e.g. low-, medium- and high-quality grading), or for screening of composition during bottling and storage. Although low accuracy and precision were obtained for the chemical parameters routinely analysed in wine, calibration models for the chemical parameters were considered acceptable for screening purposes in terms of the standard errors obtained.

Feasibility Studies↗

Image-guided in vivo dosimetry for quality assurance of IMRT treatment for prostate cancer.

PURPOSE: In external beam radiotherapy (EBRT) and especially in intensity-modulated radiotherapy (IMRT), the accuracy of the dose distribution in the patient is of utmost importance. It was investigated whether image guided in vivo dosimetry in the rectum is a reliable method for online dose verification. METHODS AND MATERIALS: Twenty-one dose measurements were performed with an ionization chamber in the rectum of 7 patients undergoing IMRT for prostate cancer. The position of the probe was determined with cone beam computed tomography (CBCT). The point of measurement was determined relative to the isocenter and relative to an anatomic reference point. The dose deviations relative to the corresponding doses in the treatment plan were calculated. With an offline CT soft-tissue match, patient positioning after ultrasound was verified. RESULTS: The mean magnitude +/- standard deviation (SD) of patient positioning errors was 3.0 +/- 2.5 mm, 5.1 +/- 4.9 mm, and 4.3 +/- 2.4 mm in the left-right, anteroposterior and craniocaudal direction. The dose deviations in points at corresponding positions relative to the isocenter were -1.4 +/- 4.9% (mean +/- SD). The mean dose deviation at corresponding anatomic positions was 6.5 +/- 21.6%. In the rare event of insufficient patient positioning, dose deviations could be >30% because of the close proximity of the probe and the posterior dose gradient. CONCLUSIONS: Image-guided dosimetry in the rectum during IMRT of the prostate is a feasible and reliable direct method for dose verification when probe position is effectively controlled.

Feasibility Studies↗