Ventricular radius reduction without resection: a computational analysis.

Abstract

Reducing a dilated left ventricle's radius by wedge resection lowers wall stress, improving performance. However, compliance falls, stroke volume improvement is limited, and, later, both functional deterioration and serious dysrhythmias are frequent. These all may result from loss of circumference, loss of contractile mass, and myocardial trauma, none of which would occur in geometric remodeling. One specific technique is bimeridianal restraint, which uses two indenting bars to remodel the left ventricle (LV) as two widely communicating "lobes" of reduced radius. Computational analysis of this technique, applied to the dilated ventricular dimensions of four pretransplant patients, showed that 20% radius reduction would be accomplished by two < or =18 mm wide bars, indenting the epicardium < or =8.3 mm with < or =6.4 mm greatest outward displacement. Projected stroke volume (SV) for the subject ventricles was then modeled and compared with projections for resected ventricles. Assuming that equally improved contractile fraction will follow equal radius reduction, however that reduction is accomplished, improvement is dramatic: if postresection remodeling SV were 1.0, 1.2, or 1.5 times baseline, then postgeometric remodeling SV would be 1.36+/-0.06, 1.66+/-0.05, or 2.14+/-0.04 times baseline, respectively. These results, preservation of contractile mass and circumferential length, complete reversibility, and minimal operative trauma, warrant study of implementing mechanical designs.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

D B Melvin. Ventricular radius reduction without resection: a computational analysis.. https://doi.org/10.1097/00002480-199905000-00011

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

3D heart model for computer simulations in cardiac surgery.

For a satisfactory computer simulation, a model, which imitates a natural situation, is needed. The Human heart is an irregular 3D object and thus difficult to reproduce. Basic data was taken from Visible Human Dataset (VHD), National Library of Medicine. The heart area was cut out of the original cross-sections and different tissues segmented. All the slices also had to be aligned to assure precise overlapping of the structures. A 3D computer heart model with the resolution of 1mm was designed. The heart model was dedicated to simulations of heat transfer during heart surgery however, it is applicable also to other medical simulations.

Cardiac Surgical Procedures↗