PubMed · 10422244
A contactless surface acoustic wave biosensor.
Abstract
Surface acoustic wave sensors operating in liquid generally cause problems resulting from wire bonding. The authors present an approach for a biosensor where the need for bonding wires is eliminated by utilizing inductive coupling of the sensor device to the RF circuitry. Protection of the electrodes from the liquid is achieved by coating the device surface with a SiO2 layer, resulting in a simplified handling of the devices. The first measurements with a sensor operating at 420 MHz are presented, demonstrating the potential of this operating principle for biosensing.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
J Freudenberg, S Schelle, K Beck, M von Schickfus, S Hunklinger. 1999-04-30. A contactless surface acoustic wave biosensor.. https://doi.org/10.1016/s0956-5663(99)00012-3