Y2K: effects on pacemaker and implantable defibrillator programmers.

Abstract

All permanent pacemakers and implantable defibrillators (PPM/ICDs) will continue to function as programmed without regard to the date in the year 2000 (Y2K). All manufacturers contacted reassured us that some of these devices incorporate a day/year clock in the circuitry; however, these are not involved in sensing or delivering programmed therapy. Some manufacturers' device programmers will roll over to the year 2000 without any problems at all, whereas others may have difficulty with date and time stamping on printed reports. We tested 14 different types of PPM/ICD programmers for Y2K compliance using 8 tests. Five of the 14 models passed each test and were labeled at our institution with a green "Y2K" sticker to identify them as Y2K compatible and needing no special attention after December 31, 1999. The most common test failed was the ability to roll the date forward from December 31, 1999, with the programmer power off. Organizations should consider testing and replacing noncompliant device programmers or placing a red sticker with "Y2K" crossed out on noncompliant pieces. The red sticker alerts the advanced practice nurse or physician to the need to confirm the appropriate date and time in the programmer after startup in the year 2000 and before interrogating or programming any PPM/ICD, to avoid inappropriate date and time stamping on printed reports from that programmer.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

D P Flynn, J P Daubert, D T Huang, C M Ocampo, E O'Gorman. Y2K: effects on pacemaker and implantable defibrillator programmers.. https://doi.org/10.1016/s0147-9563(99)70027-0

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

In the wake of hospital inquiries: impact on staff and safety.

Mishandled concerns about clinical standards resulted in whistleblowing in four Australian hospitals. Official inquiries followed with recommendations to improve patient safety. In the aftermath of the inquiries, common themes included loss of trust in management and among clinical colleagues, and loss of trust from patients and the community. Without first rebuilding trust, staff will not report mistakes or other concerns about safety. Successful implementation of patient safety procedures requires policies to stress the professional duty of staff to report concerns about colleagues when they believe there is a risk to patients.

Attitude of Health Personnel↗

Retrieving the ars moriendi tradition.

North Atlantic culture lacks a commonly shared view on dying well that helps the dying, their social environment and caregivers to determine their place and role, interpret death and deal with the process of ethical deliberation. What is lacking nowadays, however, has been part of Western culture in medieval times and was known as the ars moriendi (art of dying well) tradition. In this paper an updated version of this tradition is presented that meets the demands of present day secularized and multiform society. Five themes are central to the new art of dying: autonomy and the self, pain control and medical intervention, attachment and relations, life balance and guilt, death and afterlife. The importance of retrieving the ancient ars moriendi outreaches the boundaries of palliative medicine, since it deals with issues that play a central role in every context of medical intervention and treatment.

Attitude of Health Personnel↗