A new technique for attaching a stereotactic frame to the head: technical note.

Abstract

The aim of this study was to develop a device which allows an intermediate, painless fixation of a stereotactic frame prior to definite pin fixation. To stabilize the stereotactic frame rubber coated metal springs were used. By testing the springs on 30 volunteers with different head diameters and circumferences the optimal shape was determined. In the clinical setting 15 patients undergoing stereotactic surgery were tested, stability and patient's convenience were measured. The procedure was well tolerated and measurements revealed symmetric distances between head and frame. Therefore these metal springs are a useful accessory to the Leksell stereotactic system.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

O Heese, J Gliemroth, U Kehler, U Knopp, H Arnold. 1999. A new technique for attaching a stereotactic frame to the head: technical note.. https://doi.org/10.1055/s-2008-1053394

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Safety evaluation of large external fixation clamps and frames in a magnetic resonance environment.

Large orthopedic external fixation clamps and related components were evaluated for force, torque, and heating response when subjected to the strong electromagnetic fields of magnetic-resonance (MR) imaging devices. Forces induced by a 3-Tesla (T) MR scanner were compiled for newly designed nonmagnetic clamps and older clamps that contained ferromagnetic components. Heating trials were performed in a 1.5 and in a 3 T MR scanner with two assembled external fixation frames. Forces of the newly designed clamps were more than a factor 2 lower as the gravitational force on the device whereas, magnetic forces on the older devices showed over 10 times the force induced by earth acceleration of gravity. No torque effects could be found for the newly designed clamps. Temperature measurements at the tips of Schanz screws in the 1.5 T MR scanner showed a rise of 0.7 degrees C for a pelvic frame and of 2.1 degrees C for a diamond knee bridge frame when normalized to a specific absorption rate (SAR) of 2 W/kg. The normalized temperature increases in the 3 T MR scanner were 0.9 degrees C for the pelvic frame and 1.1 degrees C for the knee bridge frame. Large external fixation frames assembled with the newly designed clamps (390 Series Clamps), carbon fiber reinforced rods, and implant quality 316L stainless steel Schanz screws met prevailing force and torque limits when tested in a 3-T field, and demonstrated temperature increase that met IEC-60601 guidelines for extremities. The influence of frame-induced eddy currents on the risk of peripheral nerve stimulation was not investigated.

External Fixators↗