Pacemaker wires and electrodes. A follow-up study.

Abstract

In 362 patients on permanent pacing, a follow up with regard to pacemaker electrode function time and connection to pulse-generators has been carried out. In 12 patients Elema epicardial electrode were used with an average function time of 3.8 years. 232 Elema EMT 588 endocardial electrodes were used in 216 patients. Average observation time for the electrodes was 3.3 years. Sixty-eight electrodes have been followed for more than 5 years. Early electrode complications comprise 10 per cent of displacements before implantation of the pulse-generator and another 6 per cent of electrode displacements within the first 3 months after implantation. Eleven per cent of the electrodes had to be corrected due to high threshold value before implantation and another 6 per cent during the first 3 months after implanattion of pulse-generator. On hundred and fifty-five unipolar electrodes of the types Cordis, Medtronic, Elema EMT 282, and Stanium were implanted in 149 patients. Average observation time for the electrodes were 1.2 years. During the first 3 months after pacemaker implant 7 per cent of the electrodes were dislocated, 5 per cent failed due to high threshold value. Three perforations of the right ventricle occurred, without serious complications. Late complications i.e. after 3 months, for EMT 588 electrode included 4 per cent electrode dislocations, 4.5 per cent failure due to high threshold, 3 cases of wire break and 4 defects in the insulation. Many of the late electrode complications were probably caused by replacement operations for pulse-generators. The most frequent late complications for the conventional unipolar electrode was wire break which occurred in 5 cases.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

H Grendahl, E Sivertssen. 1976. Pacemaker wires and electrodes. A follow-up study.. https://doi.org/10.1111/j.0954-6820.1976.tb08372.x

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Safety of nerve conduction studies in patients with implanted cardiac devices.

Patients with implanted cardiac devices and their physicians may defer important electrodiagnostic testing because of anxiety about potential negative effects on the device. To determine the safety of routine nerve conduction studies (NCS) in this population, 10 patients with permanent dual-chamber pacemakers of various types and five patients with implanted cardiac defibrillators (ICD) underwent nerve stimulation at sites commonly used during NCS. The implanted cardiac device was interrogated before and after the study and there was continuous monitoring of the surface electrocardiogram (ECG) and atrial and ventricular electrograms. Electrical impulses generated during routine NCS were never detected by the sensing amplifier and did not affect the programmed settings of the implanted cardiac device. We conclude that routine NCS is safe in patients with implanted cardiac pacemakers with bipolar sensing configurations and defibrillators.

Arrhythmias, Cardiac↗

[Pulmonary diseases and heart function].

The most severe cardiac sequel to lung disease is the load on the right ventricle due to pulmonary hypertension with the development of a cor pulmonale. This is characterized by hypertrophy and/or dilatation of the right ventricle because of a primary impairment of lung function and/or lung structure. The most important pathomechanisms for the development of pulmonary hypertension are vessel obliteration, mechanical lesions, primary vascular or extra-vascular inflammation and hypoxic vasoconstriction. Chronic obstructive pulmonary disease (COPD) is one of the most important reasons for chronic cor pulmonale. A further very common reason is obstructive sleep apnea syndrome, especially if combined with a COPD. In this case, the prevalence of cor pulmonale can reach 80%. The development of a chronic cor pulmonale is the most striking negative prognostic factor for these patients. Only 30% of COPD patients with cor pulmonale survive longer than 5 years, and only early detection of the disturbances to respiration which might potentially lead to cor pulmonale and their subsequent therapy are able to improve the patient's prognosis. Furthermore, pulmonary diseases may also have an impact on the left heart side in terms of an impairment of left heart function or by inducing severe arrhythmias . Thus, lung diseases may have both a significant impact on right and left heart performance.

Arrhythmias, Cardiac↗