Abstract
In determining zone diameter breakpoints, the error-rate bounded method focuses directly on the observed discrepancy percentages (very major, major, and minor). These percentages, however, are quite variable due to the number of isolates investigated, the drug-specific relationship between MIC and zone diameter, the location of the isolates relative to the MIC intermediate zone, and the inherent variability of each test. To overcome potential sampling problems, a hierarchical model is proposed which explicitly accounts for each of these factors and probabilities from this model are used to determine diameter breakpoints. A simulation study is performed to demonstrate the improved consistency of this model-based procedure. Application to three published scatterplots demonstrate its interpretability advantages.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
B A Craig. 2000. Modeling approach to diameter breakpoint determination.. https://doi.org/10.1016/s0732-8893(99)00130-3