Digital mammography image simulation using Monte Carlo.
Abstract
Monte Carlo simulations of digital images of the contrast detail phantom and the ACR phantom are presented for two different x-ray digital mammography modalities: a synchrotron mammography system and a next-generation scanning slot clinical system. A combination of variance reduction methods made it possible to simulate accurate images using real pixel dimensions within reasonable computation times. The complete method of image simulation, including a simple detector response model, a simple noise model, and the incorporation of system effects (MTF), is presented. The simulated images of the phantoms show good agreement with images measured on the two systems.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
D E Peplow, K Verghese. 2000. Digital mammography image simulation using Monte Carlo.. https://doi.org/10.1118/1.598896