[Accidental decrease in the air flow during air/oxygen sevoflurane anesthesia].

Abstract

We describe our experience of an accidental decrease in the air flow during air/oxygen sevoflurane anesthesia. According to the malfunction of an air compressor system, dehydration of the compressed air was inadequate, and the wet air was delivered to the anesthesia machine. The moisture was formed at a flow control valve to cause a gradual decrease in the air flow. The low-pressure alarm did not work at that time, because the pressure of the air supply was normal. There are two types of medical air, one is produced from the ambient air by air compressor system and another is the synthetic air which is a mixture of oxygen and nitrogen. Although the synthetic air is always dry and clean, the former contains dust, bacteria, and moisture. There is a possibility that all of these particles cannot be excluded before use. Accordingly, we must check the air compressor system routinely when a source of the compressed air is used.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

M Okuyama, I Nakamura, O Kemmotsu. 2000. [Accidental decrease in the air flow during air/oxygen sevoflurane anesthesia].. https://pubmed.ncbi.nlm.nih.gov/10885247/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Flying free.

Explore the source record for details and available documents.

Air↗

Control of the redox potential by oxygen limitation improves bacterial leaching of chalcopyrite.

Shake flask and stirred tank bioleaching experiments showed that the dissolution of chalcopyrite is inhibited by ferric ion concentrations as low as 200 mg L(-1) and redox potentials >420 mV (vs. Ag/AgCl). Chemical leaching of chalcopyrite (4% suspension, surface area 2.3 m2 g(-1)) was enhanced fourfold in the presence of 0.1 M ferrous sulphate compared with 0.1 M ferric sulphate. A computer-controlled reactor was designed to function as a "potentiostat"-bioreactor by arresting the air supply to the reactor when the redox potential in solution was greater than a designated setpoint. Leaching at a low, constant redox potential (380 mV vs. Ag/AgCl) achieved final copper recoveries of 52%-61%, which was twice that achieved with a continuous supply of oxygen (<30% extraction). The bacterial populations were observed to continue growing under oxygen limitation but in a controlled manner that was found to improve chalcopyrite dissolution. As the control mechanism is easily established and is likely to decrease production cost, the use of this technology may find application in industry.

Air↗

Room scatter studies in the air kerma strength standardization of the Amersham CDCS-J-type 137Cs source: a Monte Carlo study.

Corrections for room scatter, [kSC (d, H)]RM, as a function of source-to-detector distance, d, and source-to-floor-height, H, and also departure from constant room scatter. ks1, have been computed for rooms of various sizes using Monte Carlo methods for air kerma strength standardization of the Amersham CDCS-J-type 137Cs brachytherapy source. These corrections will also be applicable to any type of 137Cs source that may be considered for standardization. It was found that, depending upon the relative position of the source with respect to the surrounding concrete scattering surfaces (side walls, floor and ceiling) and different set of d values, the assumption of constant room scatter overestimated the air kerma strength, Sk, by between 0.2% and 0.6%.

Air↗