Validation of the Yale circumferential quantification method using 201Tl and 99mTc: a phantom study.

Abstract

UNLABELLED: The Yale circumferential quantification (Yale CQ) method for quantification of SPECT images has been validated previously using empirically derived correction factors. In the present studies, the Yale CQ method was further validated using 2 SPECT gamma cameras and 2 radioisotopes. METHODS: SPECT images were acquired from cardiac phantoms with multiple fillable inserts to simulate myocardial perfusion defects of varying extents and severities. Seventy phantom configurations were created. One hundred and forty SPECT images (70 with 99mTc and 70 with 201TI) were acquired using a triple-head SPECT camera. SPECT defects were quantified using the Yale CQ method, with incorporation of 99mTc- and 201TI-derived normal databases and correction factors. RESULTS: Quantified phantom SPECT defect sizes acquired with 99mTc correlated well with actual calculated defect sizes (r = 0.96, y = 0.92x - 0.41). Bland-Altman analysis of agreement revealed strong agreement over a wide range of defect sizes, with a mean error of 1.2% and 2 SDs of 5.0%. Overall 201TI SPECT defect sizes also correlated well with actual defect sizes (r = 0.92), but there was a systematic underestimation (y = 0.72x - 0.76). Bland-Altman analysis showed underestimation over the entire range of defect sizes, with a mean error of 3.4% and 2 SDs of 7.5%. Implementation of a normal 201TI phantom database improved accuracy of quantification (r = 0.95, y = 0.87x - 1.36). The addition of 201TI-specific correction factors further improved accuracy (r= 0.94, y = 0.98x - 1.52). Reproducibility of SPECT defect sizes quantification for 99mTc using 2 gamma cameras was excellent (r = 0.98, y = 0.98x + 0.84). CONCLUSION: The Yale CQ SPECT quantification method, using the empirically derived correction factors, provides accurate and reproducible quantification of phantom defects over a wide range of defect sizes. Accurate quantification of 201TI and 99mTc SPECT defect sizes requires radiotracer-specific normal databases.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

S Kirac, F J Wackers, Y H Liu. 2000. Validation of the Yale circumferential quantification method using 201Tl and 99mTc: a phantom study.. https://pubmed.ncbi.nlm.nih.gov/10945538/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

The variation of intrinsic spatial resolution across the UFOV of scintillation cameras.

The aim of the present study was to investigate in detail the variation of the intrinsic spatial resolution across the useful field of view (UFOV) of gamma-cameras and to explore whether this variation could lead to observable effects in clinical images. Two gamma-cameras were used, without their collimators, to acquire 560 (99m)Tc point source images at different points across their UFOVs, in order to measure the intrinsic spatial resolution at each point. Possible clinical effects of the resolution variation were examined on images of a thyroid phantom using a LEHR collimator, acquired at different locations on the UFOV and at various distances from the collimator. The gamma-camera intrinsic resolution varied significantly across the UFOV, being generally lower at the central region and deteriorating at the edges. Pronounced local maxima and minima were found at points corresponding to the centers of the photomultiplier tubes (PMTs) and halfway in between. Maximum differences of more than 50% were observed between the points presenting the best and worst intrinsic resolution. Differences between neighboring points reached 17%. The effects of resolution variation were clearly observable on the thyroid phantom images. It was concluded that an appropriate correction algorithm might be necessary in order to correct for the variation of the intrinsic spatial resolution across the UFOV of gamma-cameras.

Gamma Cameras↗

Defining the lung outline from a gamma camera transmission attenuation map.

Segmentation of the lung outline from gamma camera transmission images of the thorax is useful in attenuation correction and quantitative image analysis. This paper describes and compares two threshold-based methods of segmentation. Simulated gamma camera transmission images of test objects were used to produce a knowledge base of the variation of threshold defining the lung outline with image resolution and chest wall thickness. Two segmentation techniques based on global (GT) and context-sensitive (CST) thresholds were developed and evaluated in simulated transmission images of realistic thoraces. The segmented lung volumes were compared to the true values used in the simulation. The mean distances between segmented and true lung surface were calculated. The techniques were also applied to three real human subject transmission images. The lung volumes were estimated and the segmentations were compared visually. The CST segmentation produced significantly superior segmentations than the GT technique in the simulated data. In human subjects, the GT technique underestimated volumes by 13% compared to the CST technique. It missed areas that clearly belonged to the lungs. In conclusion, both techniques segmented the lungs with reasonable accuracy and precision. The CST approach was superior, particularly in real human subject images.

Gamma Cameras↗

Performance comparison of two dual-head coincidence cameras of the first and latest generation.

This study compares the performance and image quality of two gamma camera based PET systems of the first and latest generation. We investigated two dual head coincidence gamma cameras (PRISM 2000XP and AXIS, manufactured in 1997 and 2001 by Picker/Philips) predominantly in accordance with the NEMA NU2-1994 and NU2-2001 protocols. All performance parameters except for spatial resolution and image quality were determined after measuring a standard cylinder over several half-life periods. Scatter and random fractions were evaluated with the sinogram technique. In order to determine spatial resolution and image quality we used phantoms as described in the NEMA NU2-2001 protocol. The efficiency of the new system was found to be increased. True count rate at activity levels used in clinical conditions is improved and scatter fraction is decreased substantially. Accordingly, improved spatial resolution and image quality were observed with the new system. Altogether, the AXIS represents a further approach to the performance of dedicated positron emission tomographs.

Gamma Cameras↗