PubMed · 10984234
A variable resolution x-ray detector for computed tomography: II. Imaging theory and performance.
Abstract
A computed tomography (CT) imaging technique called variable resolution x-ray (VRX) detection provides variable image resolution ranging from that of clinical body scanning (1 cy/mm) to that of microscopy (100 cy/mm). In this paper, an experimental VRX CT scanner based on a rotating subject table and an angulated storage phosphor screen detector is described and tested. The measured projection resolution of the scanner is > or = 20 lp/mm. Using this scanner, 4.8-s CT scans are made of specimens of human extremities and of in vivo hamsters. In addition, the system's projected spatial resolution is calculated to exceed 100 cy/mm for a future on-line CT scanner incorporating smaller focal spots (0.1 mm) than those currently used and a 1008-channel VRX detector with 0.6-mm cell spacing.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
F A DiBianca, P Zou, L M Jordan, J S Laughter, H D Zeman, J Sebes. 2000. A variable resolution x-ray detector for computed tomography: II. Imaging theory and performance.. https://doi.org/10.1118/1.1285831