Shockwave frequency affects fragmentation in a kidney stone model.

Abstract

PURPOSE: To study the effect of altering shockwave frequency on the efficiency of stone fragmentation using the MFL 5000 spark-gap lithotripter. MATERIALS AND METHODS: Standardized solid plaster stones, 12.0 +/- 0.5 mm in diameter, were fragmented at an energy setting of 20 kV. The shockwave frequencies tested were 60, 80, and 117 per minute. Stones were fragmented throughout the entire lifespan of the electrode, from 0 to >100% consumption, at each frequency tested. Electrode pressure output was studied for each frequency. RESULTS: A greater number of shocks was required to fragment the plaster balls at higher frequencies (regression coefficient 1.93; p < 0.003). An inverse relation was found between the number of shocks necessary to break the stones and electrode consumption (regression coefficient -2.16; p < 0.001). The analysis of delivered pressure from the electrode failed to demonstrate a linear relation with frequency (regression coefficient -0.40; p < 0.728) or consumption (regression coefficient -1.11; p < 0.158). CONCLUSIONS: The number of shocks required to fragment a stone is influenced in part by the frequency at which the shockwaves are delivered. Increasing the shockwave frequency from 60 to 117 per minute in this study caused a significant rise in the number of shocks required to break the stone. The pressure output of the electrode was similar at the frequencies tested, thus making the difference in stone fragmentation secondary to the mechanism of stone disintegration and not the function of the electrode.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

M J Weir, N Tariq, R J Honey. 2000. Shockwave frequency affects fragmentation in a kidney stone model.. https://doi.org/10.1089/08927790050152113

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Optimal designs for radiation retention with poisson correlated response.

In this paper we describe a non-linear model with correlated observations which accounts for the elimination rate of radiation in the lung of individuals who have been exposed to an accidental intake at some time. The response is then modelled as a conditional Poisson distribution. When the leak is moderate or the size of the particles is large a theoretical justification of this assumption is given and D-optimal designs are computed.

Dose-Response Relationship, Radiation↗

Nanodosemeters based on gel scintillators.

The feasibility of a nanodosemeter based on a liquid scintillator cocktail of four components (ethoxylated nonylphenol, pseudocumene, water and a lipophilic mixture) is studied. The dosemeter can work in distinct gel phases, for which the radioactive substance can be confined inside aqueous nanoscale structures of different size. For water volumes ranging 0-15%, it results in a gel with micelles of 4 nm radius. For water volumes ranging 30-50%, the resulting liquid-crystal gel contains nanostructures of approximately 20 nm radius. The low-energy electron emission arising from the decay of (3)H and (55)Fe is counted in a commercial liquid-scintillation counting spectrometer for both homogeneous and gel samples. The counting efficiency gap between the two phases is used to compute the average energy deposited inside the micelle.

Dose-Response Relationship, Radiation↗

Stereotactic radiosurgery and radiation therapy for spinal tumors.

Spinal tumors constitute 15% of all CNS neoplasms. Radiation therapy can be administered for palliation of pain and spinal cord compression. However, the amount of radiation that can be administered is often limited by the tolerance of the spinal cord, especially in cases where prior radiation therapy has been given. Stereotactic radiosurgery and radiotherapy allow the delivery of a higher dose of radiation to spinal lesions, while limiting the spinal cord dose to below the tolerance level. These are technically demanding procedures and should be performed only when proper equipment and expertise are available. Data on spinal stereotactic radiosurgery and radiotherapy have emerged in recent years. This review summarizes the clinical applications of stereotactic radiosurgery and radiotherapy for spinal tumors.

Dose-Response Relationship, Radiation↗