PubMed · 11042053
Synchronized adiabatic decoupling.
Abstract
A new decoupling scheme termed "synchronized adiabatic decoupling" is developed for use in the indirectly detected dimension. After each increment, the decoupling sequence is replaced by another one with different period T or different initial period T(ini) so that sampling always occurs at the end of a complete decoupling period. The effects of J coupling are therefore completely averaged out for all data points. As a result, all decoupling sidebands disappear and the center band increases correspondingly. Since the synchronized adiabatic decoupling does not require conventional editing techniques to cancel the sidebands, it is useful in high-field gradient-enhanced multidimensional experiments with only a single scan per increment.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
S Zhang, D G Gorenstein. 2000. Synchronized adiabatic decoupling.. https://doi.org/10.1006/jmre.2000.2183
Cite the original work for its findings. Save a collection to share your selection of sources.