PubMed · 11161185
Clinically relevant rat model for testing BOLD functional MR imaging techniques by using single-shot echo-planar imaging at 1.5 T.
Abstract
By using a 1.5-T whole-body magnetic resonance (MR) imager, a high-spatial-resolution single-shot echo-planar technique was developed to perform blood oxygen level dependent functional MR imaging of rat sensory cortex during forepaw stimulation. This technique produced cubic 1-mm(3) voxels. Signal-to-noise ratio was 140-160 (43-44 dB). Optimal effective echo time was 50 msec. This system should prove useful for developing new functional MR imaging techniques with rapid adaptation to human use.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
D W Morton, K R Maravilla, J R Meno, H R Winn. 2001. Clinically relevant rat model for testing BOLD functional MR imaging techniques by using single-shot echo-planar imaging at 1.5 T.. https://doi.org/10.1148/radiology.218.2.r01fe30598
Cite the original work for its findings. Save a collection to share your selection of sources.