A stability-indicating HPLC assay method for budesonide.

Abstract

The official (European) pharmacopeial assay for budesonide was found to be non-specific and non-stability-indicating when used to qualify several batches of pharmaceutical grade drug substance from different sources. In contrast, the most widely cited HPLC method in the literature was found to be specific and stability-indicating with respect to drug substance stored in the dry state. However, that method failed the pharmacopeia's assay system suitability requirements because of peak tailing. Moreover, it was unable to detect or resolve two major degradation products which resulted from drug storage in non-aqueous solution. A new stability-indicating HPLC method described here overcomes these problems. This method used a Hypersil C18 column with a mobile phase consisting of ethanol-acetonitrile phosphate buffer (pH 3.4; 25.6 mM) (2:30:68, v/v/v), a flow rate of 1.5 ml/min and UV detection at 240 nm. The purity of budesonide EP and its impurity profile (related substances) were tested using the new assay method, and the results compared to those from the two other methods described above. Solid-state and solution stressed stability samples were used to evaluate all methods. Using the novel method, the epimers of budesonide. their related impurities and degradation products were separated successfully. Validation studies demonstrated that the novel method possessed a linear UV response, good system precision and accuracy, high sensitivity and specificity for budesonide. The novel method will be used for future studies of budesonide's degradation kinetics.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

S Hou, M Hindle, P R Byron. 2001. A stability-indicating HPLC assay method for budesonide.. https://doi.org/10.1016/s0731-7085(00)00424-6

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

A randomized comparison of dexamethasone 0.15 mg/kg versus 0.6 mg/kg for the treatment of moderate to severe croup.

BACKGROUND: Although previous studies have shown that dexamethasone is useful in croup, the optimal dosage has not been clarified yet. The aim of this study was to investigate the effectiveness of dexamethasone 0.15 mg/kg single dose compared with the recommended dose of 0.6 mg/kg for treatment of hospitalized children with moderate to severe croup. METHOD: Forty-one children aged 6 months to 5 years admitted from March 2001 to October 2003 at the pediatric ward of Hatyai Hospital with moderate to severe croup were recruited for the study. After random allocation, each child received a single intravenous injection of either dexamethasone 0.6 mg/kg (maximum dose 12 mg) or 0.15 mg/kg (maximum dose 3 mg). All children were given a single dose of epinephrine nebulization at the beginning of the study. The croup scores were measured at 0, 1, 2, 3, 4, 6, 8, 10 and 12 h after the dexamethasone injection. RESULTS: One hour following the injection, the mean croup scores were significantly reduced from baseline values in both groups (p<0.01). There was no difference in the mean croup scores between the two groups at any time. Median times to croup scores equal or less than two in both high- and low-dose groups were 8 and 7.9 h, respectively. No patient required intubation or was removed from the study because of a rising croup score after enrollment. There was no significant adverse reaction from dexamethasone treatment in either group. CONCLUSION: Dexamethasone 0.15 and 0.6 mg/kg are equally effective in the treatment of moderate to severe croup.

Anti-Inflammatory Agents↗