Insertion gain measurements using two low-powered analogue hearing aids.

Abstract

Differences between insertion gain measurements and coupler measurements were investigated using two different hearing aids, to assess the accuracy of hearing aid fitting methods which involve the use of real-ear-to-coupler differences (RECDs). The hearing aids used were the NHS BE19 and the Lavis X11. They were allocated alternately to 80 new adult hearing aid users, and insertion gain measurement was carried out with the allocated aid on all subjects, and with both hearing aids on 36 subjects. Coupler measurements were also carried out with both hearing aids, at two different gain settings. Comparison of the coupler measurements and the insertion gain measurements suggests that the differences in insertion gain produced by different hearing aids cannot be predicted accurately from the differences in coupler gain measurements produced by the same aids. Although the use of RECDs measured using insert earphones may accurately predict the real-ear performance of some models of hearing aid, this method may not produce equally satisfactory results with other models.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

M Norman, P James. 2000. Insertion gain measurements using two low-powered analogue hearing aids.. https://doi.org/10.3109/03005364000000153

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Fitting strategies and candidature criteria for unilateral and bilateral hearing aid fittings.

In this paper, hearing aid fitting strategies are reviewed and candidature criteria for unilateral and bilateral fittings are discussed. Fitting strategies are primarily governed by the objectives of hearing aid fitting that can be set in different ways. Based on common goals, fitting strategies can be broken down into three stages: (1) basic fitting, (2) fine-tuning and (3) verification and validation. A fine-tuning concept basing on so-called meta-controllers is described and the issue of bilateral fitting strategies is discussed. In a first order approach, bilateral hearing aids can be fitted in parallel as unilateral aids, followed by bilateral fine-tuning, making sure that loudness is balanced and overall loudness is accepted. Parallel fitting of bilateral aids is more favourable than sequential fitting to avoid rejection because of a negative experience with unilateral amplification due to difficulties in disadvantageous listening situations. Candidature criteria are presented by means of a flow chart structuring the decision making process and arriving at the conclusion that bilateral fitting should be considered the normal case. If there is any doubt about the benefit of bilateral fitting, a trial with bilateral aids should be performed.

Hearing Aids↗