PubMed Health⌕ Search

PubMed · 11246942

Platelet counting by the RBC/platelet ratio method. A reference method.

Abstract

The International Council for Standardization in Haematology (ICSH) and the International Society of Laboratory Hematology (ISLH) recommend the counting of specifically labeled platelets relative to the RBCs with a fluorescence flow cytometer, together with an accurate RBC count determined with a semiautomated, single-channel aperture-impedance counter as a reference method for the enumeration of platelets. Fresh EDTA-anticoagulated venous blood specimens are measured within 4 hours of the draw. The specimen is prediluted (1:20) and the platelets labeled with two monoclonal antibodies specific to a cluster of differentiation common to all platelets. A final 1:1,000 dilution is made and at least 50,000 events with a minimum of 1,000 platelet events are counted with a flow cytometer to determine the RBC/platelet ratio. The platelet count is then calculated from this ratio and the RBC concentration of the original blood specimen.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

International Council for Standardization in Haematology Expert Panel on Cytometry, International Society of Laboratory Hematology Task Force on Platelet Counting. 2001. Platelet counting by the RBC/platelet ratio method. A reference method.. https://doi.org/10.1309/w612-myep-fa7u-8uya

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations