Increased throughput in quantitative bioanalysis using parallel-column liquid chromatography with mass spectrometric detection.

Abstract

The feasibility of quantitative bioanalysis by parallel-column liquid chromatography in conjunction with a conventional single-source electrospray mass spectrometer has been investigated using plasma samples containing a drug and its three metabolites. Within a single chromatographic run time, sample injections were made alternately onto each of two analytical columns in parallel at specified intervals, with a mass spectrometer data file opened at every injection. Thus, the mass spectrometer collected data from two sample injections into separate data files within a single chromatographic run time. Therefore, without sacrificing the chromatographic separation or the selected reaction monitoring (SRM) dwell time, the sample throughput was increased by a factor of two. Comparing the method validation results obtained using the two-column system with those obtained using the corresponding conventional single-column approach, the methods on the two systems were found to be equivalent in terms of accuracy and precision. The parallel-column system is simple and can be implemented using existing laboratory equipment with no additional capital outlays. A parallel-column system configured in this manner can be used not only for the within-a-run analysis of two samples containing two different sets of chemical entities, but also for the within-a-run analysis of two samples containing the same set of chemical entities.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

M Jemal, M Huang, Y Mao, D Whigan, M L Powell. 2001. Increased throughput in quantitative bioanalysis using parallel-column liquid chromatography with mass spectrometric detection.. https://doi.org/10.1002/rcm.330

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Sensitive and rapid liquid chromatography-tandem mass spectrometry method for the determination of stavudine in human plasma.

A sensitive method for the determination of stavudine in plasma was developed, using high-performance liquid chromatographic separation with tandem mass spectrometric detection. The samples were extracted from plasma with Waters, Sep-Pak Vac, 100 mg, tC(18) solid-phase extraction (SPE) columns. Chromatography was performed on a Supelco Discovery C(18), 5 microm, 150 x 2 mm column with a mobile phase consisting of ammonium acetate (0.01 M)-acetonitrile-methanol (800:100:100, v/v/v) at a flow-rate of 0.3 ml/min. Detection was achieved by an Applied Biosystems API 2000 mass spectrometer (LC-MS-MS) set at unit resolution in the multiple reaction monitoring mode (MRM). Atmospheric pressure chemical ionization (APCI) was used for ion production. The mean recovery for stavudine was 94% with a lower limit of quantification set at 4 ng/ml. This assay method makes use of the increased sensitivity and selectivity of mass spectrometric (MS-MS) detection to allow for a more rapid (extraction and chromatography) and selective method for the determination of stavudine in human plasma than has previously been described.

Chromatography, Liquid↗

Ion suppression effects in liquid chromatography-electrospray-ionisation transport-region collision induced dissociation mass spectrometry with different serum extraction methods for systematic toxicological analysis with mass spectra libraries.

Ion suppression effects during electrospray-ionsation mass spectrometry (ESI-MS) caused by different sample preparation procedures for serum were investigated. This topic is of importance for systematic toxicological analysis for which LC-ESI-MS has been developed with transport-region collision-induced dissociation (ECI-CID) and mass spectra library searching. With continuous postcolumn infusion of two test compounds-codeine and glafenine-the ion suppression effects of extracted biological matrix obtained after a standard liquid-liquid extraction, a mixed-mode solid-phase extraction (SPE) method, a protein precipitation method and a combination of precipitation with polymer-based mixed-mode SPE have been investigated. Extracted ion chromatograms of codeine ([M+H](+), m/z 300) and glafenine ([M-H](-), m/z 371) were used for monitoring ion suppression. Severe ion suppression effects for codeine and glafenine were detected in positive and in negative ionisation modes, respectively, in the LC-front peak after serum clean-up with SPE (acid/neutral fraction) and protein precipitation as well as with protein precipitation combined with SPE. Less ion suppression of codeine in positive mode was found with liquid-liquid extraction of serum samples. No ion suppression was detected with the second fraction of the mixed-mode SPE (using RP-C(8) and cation-exchange phase) in both ionisation modes. All suppression effects were caused by polar and unretained matrix components, which were present after extraction and/or protein precipitation. However, no specific ion suppression was seen after elution of the polar LC-front throughout the whole gradient. It could be demonstrated, that ion suppression is not generally present at any retention time when using reversed-phase HPLC with rather long gradient programs, but may play an important role in case of high-throughput LC-MS analysis, when the analyte is not separated from the LC-front, or in flow injection analysis without chromatographic separation.

Chromatography, Liquid↗

Quantification of rosuvastatin in human plasma by automated solid-phase extraction using tandem mass spectrometric detection.

An assay employing automated solid-phase extraction (SPE) followed by high-performance liquid chromatography with positive ion TurboIonspray tandem mass spectrometry (LC-MS-MS) was developed and validated for the quantification of rosuvastatin (Crestor) in human plasma. Rosuvastatin is a hydroxy-methyl glutaryl coenzyme A reductase inhibitor currently under development by AstraZeneca. The standard curve range in human plasma was 0.1-30 ng/ml with a lower limit of quantification (LLOQ) verified at 0.1 ng/ml. Inaccuracy was less than 8% and imprecision less than +/-15% at all concentration levels. There was no interference from endogenous substances. The analyte was stable in human plasma following three freeze/thaw cycles and for up to 6 months following storage at both -20 and -70 degrees C. The assay was successfully applied to the analysis of rosuvastatin in human plasma samples derived from clinical trials, allowing the pharmacokinetics of the compound to be determined.

Chromatography, Liquid↗