PubMed · 11409960
Solid phase extraction gas chromatography/electron capture detector method for the determination of organochlorine pesticides in wildlife whole blood.
Abstract
A gas chromatographic method for the analysis of 10 organochlorine pesticides in 0.5 mL of whole blood is described. Sample preparation involved an ethyl ether and hexane extraction, followed by a silica solid phase extraction cleanup. The pesticides are quantified by gas chromatography/electron capture detection. Method limits of detection ranged from 1.1 to 5.2 microg/L. The mean and standard deviation for the recovery of 10 pesticides was 97.9 +/- 5.5%. Recoveries from whole blood were comparable to recoveries from plasma. This indicates that the preparation of plasma is unnecessary for the quantification of organochlorine pesticides in blood. This approach is particularly useful as a nonlethal approach for monitoring pesticide contamination in small animals for which the volume of blood is limiting.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
S A Volz, J J Johnston, D L Griffin. 2001. Solid phase extraction gas chromatography/electron capture detector method for the determination of organochlorine pesticides in wildlife whole blood.. https://doi.org/10.1021/jf001182j