PubMed · 11419359
Ventriculo-peritoneal shunt performance under hyperbaric conditions.
Abstract
A novice scuba diver with an implanted ventriculo-peritoneal (VP) shunt inquired about the performance characteristics of his shunt while diving. A literature search revealed no information regarding shunt performance under hyperbaric conditions. The manufacturer could not certify that the shunt would function under pressure. Therefore, four VP shunts were tested according to the manufacturer's testing protocol at 1 and 4 atm abs in a multiplace hyperbaric chamber. The pressure (in mm of H2O) required to establish flow through the shunts was recorded. Trials at 1 atm abs (n = 12) and 4 atm abs (n = 12) show that all shunts performed within the pressure range specified by the manufacturer.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
E T Huang, K R Hardy, J M Stubbs, R A Lowe, S R Thom. 2000. Ventriculo-peritoneal shunt performance under hyperbaric conditions.. https://pubmed.ncbi.nlm.nih.gov/11419359/
Cite the original work for its findings. Save a collection to share your selection of sources.