PubMed · 11442103

Quality-control issues on high-resolution diagnostic monitors.

Abstract

Previous literature indicates a need for more data collection in the area of quality control of high-resolution diagnostic monitors. Throughout acceptance testing, which began in June 2000, stability of monitor calibration was analyzed. Although image quality on all monitors was found to be acceptable upon initial acceptance testing using VeriLUM software by Image Smiths, Inc (Germantown, MD), it was determined to be unacceptable during the clinical phase of acceptance testing. High-resolution monitors were evaluated for quality assurance on a weekly basis from installation through acceptance testing and beyond. During clinical utilization determination (CUD), monitor calibration was identified as a problem and the manufacturer returned and recalibrated all workstations. From that time through final acceptance testing, high-resolution monitor calibration and monitor failure rate remained a problem. The monitor vendor then returned to the site to address these areas. Monitor defocus was still noticeable and calibration checks were increased to three times per week. White and black level drift on medium-resolution monitors had been attributed to raster size settings. Measurements of white and black level at several different size settings were taken to determine the effect of size on white and black level settings. Black level remained steady with size change. White level appeared to increase by 2.0 cd/m2 for every 0.1 inches decrease in horizontal raster size. This was determined not to be the cause of the observed brightness drift. Frequency of calibration/testing is an issue in a clinical environment. The increased frequency required at our site cannot be sustained. The medical physics division cannot provide dedicated personnel to conduct the quality-assurance testing on all monitors at this interval due to other physics commitments throughout the hospital. Monitor access is also an issue due to radiologists' need to read images. Some workstations are in use 7 AM to 11 PM daily. An appropriate monitor calibration frequency must be established during acceptance testing to ensure unacceptable drift is not masked by excessive calibration frequency. Standards for acceptable black level and white level drift also need to be determined. The monitor vendor and hospital staff agree that currently, very small printed text is an acceptable method of determining monitor blur, however, a better method of determining monitor blur is being pursued. Although monitors may show acceptable quality during initial acceptance testing, they need to show sustained quality during the clinical acceptance-testing phase. Defocus, black level, and white level are image quality concerns, which need to be evaluated during the clinical phase of acceptance testing. Image quality deficiencies can have a negative impact on patient care and raise serious medical-legal concerns. The attention to quality control required of the hospital staff needs to be realistic and not have a significant impact on radiology workflow.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

L F Parr, A L Anderson, B K Glennon, P Fetherston. 2001. Quality-control issues on high-resolution diagnostic monitors.. https://doi.org/10.1007/bf03190289

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Fermentation process tracking through enhanced spectral calibration modeling.

The FDA process analytical technology (PAT) initiative will materialize in a significant increase in the number of installations of spectroscopic instrumentation. However, to attain the greatest benefit from the data generated, there is a need for calibration procedures that extract the maximum information content. For example, in fermentation processes, the interpretation of the resulting spectra is challenging as a consequence of the large number of wavelengths recorded, the underlying correlation structure that is evident between the wavelengths and the impact of the measurement environment. Approaches to the development of calibration models have been based on the application of partial least squares (PLS) either to the full spectral signature or to a subset of wavelengths. This paper presents a new approach to calibration modeling that combines a wavelength selection procedure, spectral window selection (SWS), where windows of wavelengths are automatically selected which are subsequently used as the basis of the calibration model. However, due to the non-uniqueness of the windows selected when the algorithm is executed repeatedly, multiple models are constructed and these are then combined using stacking thereby increasing the robustness of the final calibration model. The methodology is applied to data generated during the monitoring of broth concentrations in an industrial fermentation process from on-line near-infrared (NIR) and mid-infrared (MIR) spectrometers. It is shown that the proposed calibration modeling procedure outperforms traditional calibration procedures, as well as enabling the identification of the critical regions of the spectra with regard to the fermentation process.

Calibration↗

Particle sizing with a fast polar nephelometer.

We reported previously the design of a polar nephelometer that uses a rotational confocal imaging setup to enable fast scanning of the scattering phase function within a field of view of 55 degrees . The full dynamic range of the detection system can be used by increasing the signal-to-noise ratio by means of averaging successive scans. The calibration of the angular response of the instrument is achieved by obtaining the transfer function of the optical detection system using Rayleigh scatterers. Accurate particle sizing of individual polystyrene spheres (ranging from 1.5 to 9 micro m in diameter) in aqueous suspension is achieved by maximizing a correlation coefficient between precalculated tables of Mie phase functions and data obtained from the polar nephelometer. Good correlation is achieved between experimental and theoretical data, proving the functioning of the instrument as a fast and convenient particle sizer.

Calibration↗

Online standard additions calibration of transient signals for inductively coupled plasma mass spectrometry.

An online standard additions calibration method for transient signals in ICPMS is demonstrated in which a small volume of standard is injected as a spike into the sample/carrier stream, overlaying the analyte peak. This technique provides the advantages of conventional standard additions but requires only a single sample run. The method corrects for matrix effects and is suitable for transient signals in which the severity of the matrix effect changes over the analyte peak. The method uses a peak-fitting program to determine the area of the underlying peak and is shown to be effective for the determination of trace metal concentrations in both a high ionic strength matrix and in a biological matrix (urine). Eight analytes with concentrations in the range of 0.82-233.2 mug L-1 in urine were simultaneously determined using a standard spiking solution of 75 mug L-1 injected through a 100-muL loop. The measured concentrations for analytes free of spectral interferences agreed with the certified values, and the precision achieved was comparable to that achieved by the certifying agency. Using a conventional cross-flow nebulizer and Scott-type spray chamber, the accuracy obtained for online standard additions calibration was within 2%, and the precision was within 5%.

Calibration↗