Experimental flow studies in glaucoma drainage device development.

Abstract

AIMS: (I) To examine whether small holes produced by 248 nm excimer laser ablation in a polymer substrate could consistently produce a pressure drop in the desired target range (5-15 mm Hg) at physiological aqueous flow rates for use as an internal flow restrictor in a glaucoma drainage device, and (ii) to investigate whether external leakage could be reduced in comparison with conventional tube and plate glaucoma drainage devices by redesigning the exterior cross sectional shape of the portion contained within the sclerocorneal tunnel. METHODS: Single holes with target diameters of 10 microm, 15 microm, 20 microm, and 25 microm were drilled using a 248 nm excimer laser in sample discs (n=6 at each diameter) punched from a 75 microm thick polyimide sheet. Sample discs were tested in a flow rig designed to measure the pressure drop across the discs. Using filtered, degassed water at a flow rate of 1.4 microl/min repeated flow measurements were taken (n=6) for each disc. After flow testing, all discs were imaged using a scanning electron microscope and the dimensions of each hole were derived using image analysis software. In the external leakage study, corneoscleral buttons (n=13) were prepared from cadaver pig eyes and mounted on an artificial anterior chamber infused with Tyrode solution. After the pressure had stabilised, standard occluded silicone tube implants were inserted through 23 gauge needle stab incisions at the limbus. These were compared against prototype PMMA implants with a novel shape profile inserted through 1.15 mm width microvitreoretinal (MVR) stab incisions at the limbus. The infusion rate was maintained and a second pressure measurement was taken when the pressure had stabilised. The difference between the first and second pressure measurement was then compared, as an index of external leakage. RESULTS: Ablated tubes were found to have a near perfect circular outline on both the entry and exit side. The observed pressure drops across the ablated sample discs at each target diameter were as follows: 10 microm, mean 25.66 (SD 4.9) mm Hg; 15 microm, 6.7 (1.15); 20 microm, 1.66 (1.07); and 25 microm, <0.1 mm Hg. A strong correlation was observed between observed pressure drops and those predicted by Poiseuille's formula (R(2) =0.996). Target ablations of 15 microm diameter produced tubes that consistently achieved a pressure drop within the desired range (5-15 mm Hg). In the external leakage study, preinsertion pressures (mm Hg; mean (SD)) were 19.00 (4.3) (conventional method) and 20.00 (3.9) (new technique with PMMA prototypes). Post-insertion pressures were significantly reduced (10.40 (7.7); p<0.01) for the conventional technique and were essentially unchanged for the new technique (18.80 (4.9); p>0.1). CONCLUSIONS: It was shown that it is possible, in principle, to control the dimensions of a manufactured tubular lumen in a glaucoma drainage device accurately enough to provide consistent protection from hypotony in the early period after glaucoma filtration surgery. By redesigning the external profile of glaucoma drainage device and incision technique, it was also shown that it is possible to eliminate uncontrolled external leakage.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

AGFID project team. 2001. Experimental flow studies in glaucoma drainage device development.. https://doi.org/10.1136/bjo.85.10.1231

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Protocol for Detecting and Sequencing Chikungunya Virus from Field-Collected Mosquitoes.

Arboviral diseases represent a major public health challenge, especially in tropical regions where environmental conditions may favor the proliferation and spread of mosquito vectors. Thus, early and accurate detection of chikungunya virus (CHIKV) in mosquito populations can be a valuable tool for effective surveillance of circulating variants and for identifying new viral introductions. Given the challenges of detecting arboviruses in field-captured mosquitoes, we describe an integrated workflow for CHIKV molecular detection and whole-genome sequencing. This protocol includes mosquito homogenization using a bead-based mechanical disruptor, RNA extraction using TRIzol reagent with minor modifications, molecular screening using CHIKV-specific RT-qPCR, and whole-genome amplification followed by sequencing on Illumina platforms. Despite the protocol being optimized for individual mosquitoes, it results in high-quality RNA suitable for both entomological surveillance and genomic analysis. As this protocol allows recovery of complete CHIKV genomes from mosquito specimens, it can serve as a basis for genomic epidemiology studies, enabling monitoring of viral diversity and lineage dynamics, and facilitating early detection of emerging variants to support timely and targeted public health interventions in endemic and at-risk regions.

Animals↗

Genomic Profiling of Chromatin State Using CUT&Tag.

Alterations in chromatin state, mediated through histone modifications and the incorporation of histone variants, are fundamental to establishing transcriptional networks and cell identity. Recent advances in low-input epigenome profiling methods, such as CUT&Tag and CUT&RUN, have enabled the study of chromatin states from very limited starting materials. In this chapter, we describe procedures for generating CUT&Tag libraries to profile histone modifications and histone variants in early-developing zebrafish embryos.

Animals↗

Relaxin-2: Shaping the Proteomic Landscape of Skeletal Muscle Physiology, Glucose Trafficking, and Mitochondrial Function in Rat.

Relaxin-2 is a hormone with robust beneficial effects on the heart and blood vessels and potential as a therapy for cardiovascular (CV) disease. Considering the interorgan communication between skeletal muscle and heart, and the relation between muscle quality/composition and CV events, we hypothesize that relaxin-2 may regulate skeletal muscle physiology and metabolism. We aim to evaluate the impact of relaxin-2 on the proteome of skeletal muscle from healthy Sprague-Dawley rats. Animals were treated with 0.4&#x2009;mg/kg/day of serelaxin (recombinant form of human relaxin-2) or vehicle (PBS) for 2&#x2009;weeks employing subcutaneous osmotic minipumps. Skeletal muscle protein identification and quantification were performed by LC-MS/MS using a Data-Independent Acquisition (DIA)-Sequential Window Acquisition of All Theoretical Fragment Ion Spectra (SWATH) method. SWATH/MS quantitative analysis identified that relaxin-2 significantly decreased 95 proteins and significantly increased 32 proteins in rat skeletal muscle when compared to control rats. From these, 34 proteins were associated with muscle function, myogenesis, muscle differentiation and/or regeneration, 20 are mitochondrial proteins (six from the complexes of the electron transport chain), and 10 proteins participate in glucose metabolism. Qualitative data-dependent workflow analysis identified 35 proteins exclusive to the skeletal muscle of the relaxin-2-treated group: eight proteins related to processes of skeletal muscle function (size, ion homeostasis or organization of caveolae structures and cytoskeleton) and myogenesis, and two proteins involved in muscle differentiation. Our work highlighted for the first time the role of relaxin-2 in crucial processes of muscle physiology and energetic metabolism, which could influence several processes involved in myopathy and CV.

Animals↗