Lab-on-a-chip sample preparation using laminar fluid diffusion interfaces--computational fluid dynamics model results and fluidic verification experiments.

Abstract

Microfluidic structures for the generation of laminar fluid diffusion interfaces (LFDIs) for sample preparation and analysis are discussed. Experimental data and the results of fluid modeling are shown. LFDIs are generated when two or more streams flow in parallel in a single microfluidic structure without any mixing of the fluids other than by diffusion of particles across the diffusion interface. It has been shown that such structures can be used for diffusion-based separation and detection applications. The method has been applied to DNA desalting, the extraction of small proteins from whole blood samples, and the detection of various constituents in whole blood, among other examples. In this paper the design and manufacture of self-contained microfluidic cartridges for the extraction of small molecules from a mixture of small and large molecules by diffusion is demonstrated. The cards are operated without any external instrumentation, and use hydrostatic pressure as the driving force. The performance of the cartridges is illustrated by separating fluorescein from a mixture of fluorescein and dextran of molecular weight 2 x 10(6). In a single pass, 98.6% of dextran was retained in the product whereas 43.1% of fluorescein was removed. The method is adjustable for different separation requirements, and computational fluid dynamics (CFD) models are shown that demonstrate the tuning of various microfluidic parameters to optimize separation performance. Other applications of LFDIs for establishment of stable concentration gradients, and the exposure of chemical constituents or biological particles to these concentration gradients are shown qualitatively. Microfluidic chips have been designed for high-throughput screening applications that enable the uniform and controlled exposure of cells to lysing agents, thus enabling the differentiation of cells by their sensitivity to specific agents in an on-chip cytometer coupled directly to the lysing structure.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

B H Weigl, R L Bardell, N Kesler, C J Morris. 2001. Lab-on-a-chip sample preparation using laminar fluid diffusion interfaces--computational fluid dynamics model results and fluidic verification experiments.. https://doi.org/10.1007/s002160100997

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Extended application of gel-permeation chromatography by spin column.

Separation of small volumes of proteins from unbound ligands or reequilibration with buffer by passing through a 1-ml Sephadex G-50 column under mild centrifugal force is a popular technique. Here it has been demonstrated that other Sephadex matrix could similarly be used for complete or partial separation of protein molecules. Proteins to be eluted at void volume are recovered near quantitatively, while others are partly or almost completely retained depending on molecular size. Calibration curves using standard proteins of Mw 12.5 to 440 kDa with Sephadex G-50-G-200 representing recovery versus molecular weight show profiles as expected from the fractionation ranges of the column matrix. The procedure may be applied to follow protein association-dissociation reactions if the molecular weights of the species concerned are known and a proper matrix exists for separating them. Equilibrium unfolding transitions constructed with model proteins in presence of 0-8M urea using recovery as an index correspond to profiles obtained from other physical measurements. This may be a convenient approach to follow change of protein hydrodynamic volume quickly when a parallel methodology is not readily available.

Chemistry Techniques, Analytical↗

Analysis of nitrated proteins by nitrotyrosine-specific affinity probes and mass spectrometry.

Tyrosine nitration is a well-established protein modification that occurs in disease states associated with oxidative stress and increased nitric oxide synthase activity. Nitration of specific tyrosine residues has been reported to affect protein structure and function, suggesting that 3-nitrotyrosine formation may not only be a disease marker but may also be involved in the pathogenesis of some diseases and in normal regulatory processes. It has been, however, difficult to identify sites of nitration. We describe a method that combines specific isolation of nitrated proteins with mass spectrometric determination of the amino acid sequence and the site of nitration of individual proteins. A complex protein mixture, e.g., serum or cell lysate, was enriched for nitrotyrosine-containing proteins by immunoprecipitation with antinitrotyrosine antibodies. The nitrotyrosines were then reduced to aminotyrosines with a strong reducing agent in parallel in-gel and in-solution procedures. Using nitrated human serum albumin as a model, we reduced the disulfide bonds with dithiothreitol and alkylated the free sulfhydryl groups with iodoacetamide. The nitrotyrosines were next reduced to aminotyrosines with sodium dithionite, and-at pH 5.0-cleavable biotin tags were selectively attached to the aminotyrosines and the albumin was then digested with trypsin. The biotinylated tryptic peptides were purified on a streptavidin affinity column and identified by mass spectrometry. We have also purified nitrated human serum albumin from an enriched sample of SJL mouse plasma and confirmed its identity by peptide mass fingerprinting and MASCOT.

Chemistry Techniques, Analytical↗

Robust microsatellite instability (MSI) analysis by denaturing high-performance liquid chromatography (DHPLC).

Microsatellite instability (MSI) plays an important biological role in various types of cancers, and especially in colorectal cancers. This study aimed to develop a simple, efficient, new method for robust MSI analysis. DNA was extracted from 175 (105 proximal colon and 70 distal colorectal) cancer samples and matched normal tissues, and five Bethesda microsatellite markers (BAT-25, BAT-26, D5S346, D2S123, and D17S250) were examined for MSI by denaturing high-performance liquid chromatography (DHPLC) analysis at a temperature of 50 degrees C and a flow rate of 0.9 ml/min. It took just 9 min per PCR product to determine MSI or microsatellite stability (MSS) using the new protocol. The DHPLC results were confirmed with conventional gel-based electrophoresis and capillary-based sequencing method. Of 175 samples, 45 (26%) showed high microsatellite instability (MSI-H), 12 (7%) showed low microsatellite instability (MSI-L), and 118 (67%) showed MSS. All MSI samples were deletion mutants and all 12 MSI-L cases had MSI in dinucleotide markers (D5S346, D2S123, and D17S250). MSI was significantly associated with proximal colon cancers ( p<0.0001), as previously reported. The MSI-H tumors were also associated with tumor node metastasis (TNM) I/II stages ( p=0.05) and high-grade tumors ( p<0.01). Here, we propose a DHPLC-based method as an alternative for MSI analysis.