PubMed · 11694283
Multi-analyte analysis of biological fluids with a recycling immunoaffinity column array.
Abstract
A system for isolating and measuring up to 30 analytes in a single biological sample is described. The system is based on recycling a pre-labeled sample through an array of capillary immunoaffinity columns, each packed with glass beads, coated with a different antibody, thus enabling each column to isolate and extract a single analyte. Detection of the bound analytes is achieved by laser-induced fluorescence (LIF), using a laboratory-built scanning detector coupled to a fiber-optic spectrometer. The array can be regenerated up to 200 times, provided a suitable temperature is maintained. The individual immunoaffinity columns are able to bind between 2.9 and 3.6 ng of analyte, depending upon the individual column, with lower limits of detection (LOD) in the order of 1.6-2.8 pg/ml. The inter- and intra-assay coefficients of variation (CV) for all 30 columns in the array were less than 6.03+/-0.33 at analyte concentrations of 100 pg/ml. Comparison to standard enzyme-immunoassays demonstrated r(2) values in the range of 0.9151-0.9855 when analyzed by least-squares linear regression.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
T M Phillips. 2001-10-30. Multi-analyte analysis of biological fluids with a recycling immunoaffinity column array.. https://doi.org/10.1016/s0165-022x(01)00202-0
Cite the original work for its findings. Save a collection to share your selection of sources.