PubMed · 11750751
Multigate doppler signal analysis using 3-D regularized long AR modelling.
Abstract
Autoregressive (AR) modelling has already been proposed as an alternative to fast Fourier transform to process ultrasound (US) Doppler signals. Previous works introduced long AR models, set up under a regularization framework. The latter may be in 1-D (frequency) or 2-D (frequency and space or time). This study generalizes the spectrum regularization in the three dimensions frequency, space and time. The problem of the penalization function is addressed, and a new convex solution is proposed, taking into account possible nonstationarity of the Doppler signal. The parameter tuning is based on simulations using a standard Doppler signal model. The first results show that this processing improves the spectral estimation, and is well suited to flow interpretation.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
C Berthomier, A Herment, J F Giovannelli, G Guidi, L Pourcelot, B Diebold. 2001. Multigate doppler signal analysis using 3-D regularized long AR modelling.. https://doi.org/10.1016/s0301-5629(01)00446-x
Cite the original work for its findings. Save a collection to share your selection of sources.