PubMed Health⌕ Search

PubMed · 12457339

Technical complication in deep brain stimulation.

Abstract

With a growing number of patients treated with deep brain stimulation (DBS) operations for both hardware-related complications and routine replacements of impulse generators will be performed more frequently. Failure of DBS systems have to be analyzed thoroughly as this thwarts the enormous efforts required for proper electrode implantation and operative revisions increase the morbidity associated with DBS. A female patient implanted with DBS electrodes for advanced Parkinson's disease presented with straining of the right extension lead and deteriorating gait because of electrode migration. This was due to a malpositioned set screw connector adapting the electrode lead to the extension wire which had been placed below the mastoid process. Following surgical revision with implantation of a new electrode into the STN, electrode dislocation recurred requiring another surgical revision. This was due to renewed connector migration from its parietal position into the cervical region. Straining of extension leads should be recognized as a warning sign for (imminent) electrode dislocation or lead fracture. This may just be the case with connectors located below the mastoid process or in the cervical region, a risk which appears to be increased further with reduced-length extensions. Renewed dislocation of revised extensions may be prevented by securing the position of the connector (e.g. with manipulates).

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

W Hamel, B Schrader, D Weinert, J Herzog, D Müller, G Deuschl, J Volkmann, H M Mehdorn. 2002. Technical complication in deep brain stimulation.. https://doi.org/10.1055/s-2002-35822

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

The development of a knee locker with closed-loop functional electrical stimulation (FES) for hemiplegia in gait training.

PURPOSE: A knee locker with closed-loop functional electrical stimulation (FES) system has been developed to prevent the quadriceps weakness and the drop-foot of the hemiplegia during gait training. METHOD: The FES system is triggered by a footswitch on the heel of the disabled foot to stimulate the tibialis anterior muscle for dorsi-flexion and to turn-off the knee locker in the swing phase through the main controller. Besides, the footswitch on the heel of the affected-side can be used to stimulate the quadriceps and to turn-on the knee locker for quadriceps weakness in the stance phase. RESULTS: It is revealed that the mean velocity, cadence, stride length, active ankle motion range, and functional ambulation category (FAC) have improved significantly from 0.15 +/- 0.04 m sec( - 1), 43.3 +/- 15.4 steps min( - 1), 0.36 +/- 0.11 m, 15 degrees, level 2 to 0.43 +/- 0.21 m sec( - 1), 69.4 +/- 19.1 steps min( - 1), 0.73 +/- 0.17 m, 40 degrees, level 4 respectively for the patient. A paired t-test indicated that differences in the electromyography (EMG) of the tibialis anterior and the quadriceps muscles between patient's disabled (affected-side) foot and normal (unaffected-side) foot are not significant (p > 0.05) after 16 weeks of training. CONCLUSIONS: It is concluded that this new knee locker with closed-loop FES system is capable of providing the hemiplegia with restoration to regular walking after appropriate gait training.

Electric Stimulation Therapy↗

Transcranial magnetic stimulation: studying motor neurophysiology of psychiatric disorders.

RATIONALE: Transcranial magnetic stimulation (TMS) is a noninvasive tool that directly stimulates cortical neurons by inducing magnetic and secondary electric fields. Traditionally TMS has been used to study the motor neurophysiology of healthy subjects and those with neurological disorders. OBJECTIVE: Given the known motor dysfunctions in many psychiatric disorders supplemental usage of TMS to study the underlying pathophysiology of certain psychiatric disorders and to assess treatment outcomes is underway. Such studies include examination of motor neuronal membrane, corticospinal and intracortical excitability. Our objective is to overview the past findings. METHODS: We review the past literature that used TMS as an assessment tool in psychiatric disorders such as schizophrenia, mood disorders, Tourette's syndrome, obsessive-compulsive disorder, attention-deficit hyperactivity disorder, and substance abuse. RESULTS: While the findings are still preliminary due to small sample-size, inconsistent patient population (diagnosis, medication), differences in methodology between research groups, studies restricted to the motor region and possible lack of sensitivity and specificity, the studies are yielding interesting results which could potentially lead to trait- and state-markers of psychiatric disorders. CONCLUSIONS: Future studies using TMS alone or in combination with other neuroimaging techniques promise to further expand the application of TMS from studies of motor excitability to higher cognitive functions.

Electric Stimulation Therapy↗