MLC quality assurance techniques for IMRT applications.

Abstract

Intensity modulated radiotherapy (IMRT) requires extensive knowledge of multileaf collimator (MLC) leaf positioning accuracy, precision, and long-term reproducibility. We have developed a technique to efficiently measure the absolute position of each MLC leaf, over the range of leaf positions utilized in IMRT, based on dosimetric information. A single radiographic film was exposed to 6 MV x-rays for twelve exposures: one open field with a radio-opaque marker tray present, and eleven fields (1 x 28 cm strips via 1 cm gaps between opposed leaf pairs) separated by 2 cm center to center. The process was repeated while varying direction of leaf travel; each film was digitized using a commercial film dosimetry system. The digital images were manipulated to remove translation and rotation of the film data with respect to the collimator coordinate system by extraction of radiation dose profiles perpendicular to the MLC leaf motion and measuring the center of the x-ray leakage between leaves. Radiation dose profiles in the direction of leaf motion were acquired through the center of each leaf pair (leaves 2-28), which provided leaf position information every 2 cm with 0.2 mm precision. Nine separate leaf reproducibility studies over a 90 day period which evaluated 600 measurement points on each film show 0.3 mm precision for 95% confidence, while hysteresis studies show 0.5 mm precision. Absolute leaf position error measurements demonstrated a radial dependence, with a maximum of 1.5 mm at 16.4 cm from central axis, due to rotational error at calibration. Recalibration of the MLC leaves based utilizing this tool yields absolute leaf position measurements where 91.5% of all leaves/positions were within 0.5 mm, with a mean error of 0.1 mm and a maximum error less than 1.0 mm.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

J E Bayouth, D Wendt, S M Morrill. 2003. MLC quality assurance techniques for IMRT applications.. https://doi.org/10.1118/1.1564091

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Dual-trap technique for reduction of low-frequency noise in force measuring optical tweezers.

High-resolution long-time force measurements by optical tweezers are often limited by low-frequency (1/f) noise. A dual-trap technique is presented that can reduce such noise in the force signal. It incorporates a second trap (a reference trap) that probes the noise in the system and it is based upon the assumption that the low-frequency parts of the noise from the two traps are correlated. A subtraction of the low-frequency signal from the reference trap from the signal from the force measuring trap will therefore yield a net signal that is significantly less influenced by noise. It is shown that this dual-trap technique can reduce the noise in the force signal up to 60% depending on detection bandwidth.

Equipment Design↗

Optimization of doses received by the hospital staff and the members of the family of patients undergoing 111In-DTPA-D-Phe1-Octreotide therapy.

According to the Euratom Directives (96/29, 97/43), the doses received by the workers as well as the family of patients and third persons during medical exposures, should conform to the dose constraint levels (DCLs), established by the authorities for each group in the context of optimisation. This study deals with the implementation of a radiation protection protocol, concerning the aforementioned group members for patients undergoing treatment with 111In-DTPA-D-Phe1-Octreotide, after intra-arterial infusion. It is shown that by applying this protocol the annual doses to the medical and technical staff are considerably reduced and remain below the established DCLs. Following the post-release behaviour instructions given to the patient, doses to the family and third persons may be kept lower than the corresponding DCLs provided by the National Regulations.

Equipment Design↗

In vivo myograph measurement of muscle contraction at optimal length.

BACKGROUND: Current devices for measuring muscle contraction in vivo have limited accuracy in establishing and re-establishing the optimum muscle length. They are variable in the reproducibility to determine the muscle contraction at this length, and often do not maintain precise conditions during the examination. Consequently, for clinical testing only semi-quantitative methods have been used. METHODS: We present a newly developed myograph, an accurate measuring device for muscle contraction, consisting of three elements. Firstly, an element for adjusting the axle of the device and the physiological axis of muscle contraction; secondly, an element to accurately position and reposition the extremity of the muscle; and thirdly, an element for the progressive pre-stretching and isometric locking of the target muscle. Thus it is possible to examine individual in vivo muscles in every pre-stretched, specified position, to maintain constant muscle-length conditions, and to accurately re-establish the conditions of the measurement process at later sessions. RESULTS: In a sequence of experiments the force of contraction of the muscle at differing stretching lengths were recorded and the forces determined. The optimum muscle length for maximal force of contraction was established. In a following sequence of experiments with smaller graduations around this optimal stretching length an increasingly accurate optimum muscle length for maximal force of contraction was determined. This optimum length was also accurately re-established at later sessions. CONCLUSION: We have introduced a new technical solution for valid, reproducible in vivo force measurements on every possible point of the stretching curve. Thus it should be possible to study the muscle contraction in vivo to the same level of accuracy as is achieved in tests with in vitro organ preparations.

Equipment Design↗