PubMed · 1293439
Pupillometer for clinical applications using dual 256-element linear CCD arrays.
Abstract
A new pupillometer for measuring the size of the pupil of the human eye is described. It uses two inexpensive 256-element charge coupled device (CCD) arrays in an orthogonal configuration. The system is controlled and data are analysed by means of a Z80A microcomputer. Software allows the position and diameter of the pupil to be displayed on a monochrome video monitor. Infra-red diodes irradiate the eye so that measurements can be made in darkness. The accuracy is better than +/- 0.1 mm, over the range 3.2-9.5 mm. Dynamic measurements of the pupil light reflex response can be made with sampling periods as short as 20 ms.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
D P Jones, S J Harris, S R Scicinski. 1992. Pupillometer for clinical applications using dual 256-element linear CCD arrays.. https://doi.org/10.1007/bf02457826