An improved ELISA with linear sweep voltammetry detection.
Abstract
An improved ELISA combined with linear sweep voltammetry detection of p-nitrophenol generated by an enzyme has been investigated in this study. p-nitrophenol, produced from alkaline phosphatase catalysing p-nitrophenyl phosphate, yielded an oxidative peak at 1.06 V (vs. Ag/AgCl) with a wax-impregnated tubular graphite anode. Without separation, the small three-electrode system was directly inserted in the well of an ELISA plate for detection. The detection limit for p-nitrophenol was 1 x 10(-6) M, lower than that obtained by measuring the absorbance of p-nitrophenol. The feasibility of utilizing linear sweep voltammetry as a detection scheme was demonstrated by determining metallothionein, granulocyte-colony stimulating factor and Xenopus laevis keratin using the above new system. The method was simple, reproducible and much more sensitive than traditional spectrophotometry.