PubMed · 1399057
Variables affecting electronic root canal measurement.
Abstract
This study was conducted in two parts. In the first part, 20 single-rooted teeth that had been scheduled for extraction were investigated. The electronic root canal lengths were measured in vivo with a Dental Sono-Explorer type Y-III, and the actual canal lengths were measured after extraction of the teeth. The rate of agreement of the two measurements was 77.5% within a range of +/- 0.5 mm, while it was 100% at +/- 2.0 mm, which is acceptable clinically. In the second part, there were 19 simulated canals whose lengths and apical foramen sizes were known beforehand. Experiments revealed a negative correlation between the areas of the apical foramina and the difference between the electronic and the actual root canal lengths. This relationship was shown by the linear regression equation: ŷ = 0.6-1.6x. With the exception of the smallest areas of foramina, electronic root canal length measurements were less than the actual lengths.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Y N Wu, J N Shi, L Z Huang, Y Y Xu. 1992. Variables affecting electronic root canal measurement.. https://doi.org/10.1111/j.1365-2591.1992.tb00741.x