PubMed · 1451596
Drop-delay measurement using enzyme-coated particles.
Abstract
A simple technique is suggested for the measurement of drop delay for flow sorting. While the flow cytometer was set to sort a fixed number of particles, the drop-delay setting was changed step by step, and at each step the HRP-coupled particles were sorted into a well of an immunoassay strip. Then the HRP activity of the sorted samples was revealed by routine methods. The maximum level of the enzyme activity shows the proper drop-delay setting. Determination of the drop-delay setting takes only a few minutes. The technique is independent of the type of flow cytometer and does not require any additional equipment.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Y A Lazebnik, A I Poletaev, V V Zenin. 1992. Drop-delay measurement using enzyme-coated particles.. https://doi.org/10.1002/cyto.990130614. https://pubmed.health/record/pubmed%3A1451596/drop-delay-measurement-using-enzyme-coated-particles
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite PubMed Health by Indiciae: BibTeX · RIS.