PubMed · 1461209
A rapid method for electron beam energy check.
Abstract
Assessment of electron beam energy and its long term stability is part of standard quality assurance practice in radiation oncology. Conventional depth-ionization or depth-film density measurements are time consuming both in terms of data acquisition and analysis. A procedure is described utilizing ionization measurements at two energy specific depths. It is based on a linear relationship between electron beam energy and its practical range. Energy shifts within the range covered by the two measurement depths are easily resolved. Within a range of +/- 0.50 MeV (+/- 1.30 MeV) around the established mean incident energy of 5.48 MeV (20.39 MeV), the method accuracy is better than 0.10 MeV.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
V J Pisciotta, X Wu, J A Fiedler, C F Serago, M L Calish-Watzich, J G Schwade, P V Houdek. A rapid method for electron beam energy check.. https://doi.org/10.1118/1.596801
Cite the original work for its findings. Save a collection to share your selection of sources.