PubMed · 1487979
Haemoglobin A2 measurement using high performance liquid chromatography.
Abstract
Haemoglobin A2 levels were assessed using a modular high performance liquid chromatography (HPLC) system with a protocol designed for the measurement of haemoglobin A1C. There was good correlation (r = 0.96; P < 0.001) between this technique and the International Committee for Standardisation in Haematology (ICSH) recommended method of microchromatography. The range of haemoglobin A2 was found to be 2.3-3.2% in apparently normal individuals, and 4.0-6.7% in those with beta-thalassaemia trait. The HPLC system produced reliable results quickly for haemoglobin A2 with no alteration to the protocol used for measuring haemoglobin A1C.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
R Deacon-Smith, I Lord. 1992. Haemoglobin A2 measurement using high performance liquid chromatography.. https://pubmed.ncbi.nlm.nih.gov/1487979/