PubMed Health⌕ Search

PubMed · 15074512

Ultrasound attenuation imaging using compound acquisition and processing.

Abstract

A method that combines both spatial and frequency compounding is described for measuring attenuation in tissue. The technique applies a reference phantom to account for imaging system dependencies of echo signals. Emphasis is given to local attenuation estimates, to reduce the variance of the attenuation measurements over small regions of interest (ROI) and to enable coarse attenuation imaging. Experiments using a uniform phantom show that the standard deviation of local attenuation estimates within a ROI drops when greater degrees of compounding are applied. Attenuation images of a specially designed phantom containing inclusions with attenuation contrast illustrate the accuracy and precision of the technique.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Haifeng Tu, Tomy Varghese, Ernest L Madsen, Quan Chen, James A Zagzebski. 2003. Ultrasound attenuation imaging using compound acquisition and processing.. https://doi.org/10.1177/016173460302500403

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

[Two approximations to obtain the collimator scatter factor (Sc) for MLC irregular fields].

The collimator scatter factor (S(c)(MLC)) at MLC irregular fields for high-energy X-ray irradiation is generally assumed to be equal to the jaw collimator scatter factor (S(c)(jaw)) of the square field equivalent to the rectangular field produced using pairs of jaw collimators. However, this assumption becomes strained as the ratio of the MLC equivalent square field side to the jaw collimator equivalent square field side decreases. In this study, for 4 MV and 10 MV X-rays, the collimator scatter factor (S(c)(MLC)) for an MLC irregular field could be evaluated with a high degree of accuracy using the MLC irregular correction (F(MIC)) factor or the jaw collimator correction (F(JCC)) factor.

Phantoms, Imaging↗

Dose reduction in multidetector CT of the urinary tract. Studies in a phantom model.

A novel ureter phantom was developed for investigations of image quality and dose in CT urography. The ureter phantom consisted of a water box (14 cm x 32 cm x 42 cm) with five parallel plastic tubes (diameter 2.7 mm) filled with different concentrations of contrast media (1.88-30 mg iodine/ml). CT density of the tubes and noise of the surrounding water were determined using two multidetector scanners (Philips MX8000 with four rows, Siemens Sensation 16 with 16 rows) with varying tube current-time product (15-100 mAs per slice), voltage (90 kV, 100 kV, 120 kV), pitch (0.875-1.75), and slice thickness (1 mm, 2 mm, 3.2 mm). Contrast-to-noise ratio as a parameter of image quality was correlated with dose (CTDI) and was compared with image evaluation by two radiologists. The CT densities of different concentrations of contrast media and contrast-to-noise ratio were significantly higher when low voltages (90 kV versus 120 kV, 100 kV versus 120 kV) were applied. Smaller slice thickness (1 mm versus 2 mm) did not change CT density but decreased contrast-to-noise ratio due to increased noise. Contrast phantom studies showed favourable effects of low tube voltage on image quality in the low dose range. This may facilitate substantial dose reduction in CT urography.

Phantoms, Imaging↗

Development of a simple control system for uniform proton dose distribution in a dual-ring double scattering method.

In proton radiotherapy with high focusing of irradiation on the tumour, it is important to obtain treatment beams with a highly uniform dose distribution. Uniform dose distribution in the clinical irradiation field can be obtained by the dual-ring double scattering method. This method is superior to the wobbler method, which uses electromagnetic deflection of the proton beams, because of the absence of the temporal structure of irradiation distribution. However, in the dual-ring double scattering method the condition of incident proton beams entering the scatter, especially the accuracy of the position of the incident proton beams with respect to the scatter, markedly affects the uniformity of the beam distribution in the irradiation field. In this study, to ensure the uniformity of dose distribution during treatment, we developed a control system equipped with an automatic fine adjustment of the beam axis and a mechanism for moving the second dual-ring scatter of the double scatters to the optimal position. Using this system, we achieved uniform dose distribution in the irradiation field during proton radiotherapy, with symmetry within +/-1% and flatness within 2%.

Phantoms, Imaging↗