PubMed Health⌕ Search

PubMed · 15123929

Image guidance: fluoroscopic navigation.

Abstract

Computer-assisted orthopaedic surgery slowly is making its way into routine orthopaedic practice. Orthopaedic trauma has long been identified as a potential impact area of this new technology. Early experience with three-dimensional (3D) image-guided surgery was promising, but this particular technique was limited by the inability to update the 3D computer model in the operating room after fracture reduction maneuvers or implant placement. Virtual fluoroscopy, or fluoroscopic navigation, became available in 1999 and has proven to be a more versatile technology for fracture treatment. Fluoroscopic navigation systems allow the surgeon to store multiple intraoperative fluoroscopic images on a computer workstation; the position of special optically-tracked surgical instruments or implants then may be virtually overlaid onto the stored images in multiple planes during implant placement. The ability to update images after fracture manipulation now has expanded the application of computer-assisted surgery to any procedures that traditionally have relied on intraoperative C-arm use. In selected applications, this technology has been shown to decrease operative time and intraoperative radiation exposure. The advantages of the new technique of fluoroscopic navigation and its current use in trauma applications will be discussed.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

David M Kahler. 2004. Image guidance: fluoroscopic navigation.. https://pubmed.ncbi.nlm.nih.gov/15123929/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Population and patient-specific target margins for 4D adaptive radiotherapy to account for intra- and inter-fraction variation in lung tumour position.

In this work, five 4D image-guidance strategies (two population, an offline adaptive and two online strategies) were evaluated that compensated for both inter- and intra-fraction variability such as changes to the baseline tumour position and respiratory pattern. None of the strategies required active motion compensation such as gating or tracking; all strategies simulated a free-breathing-based treatment technique. Online kilovoltage fluoroscopy was acquired for eight patients with lung tumours, and used to construct inter- and intra-fraction tumour position variability models. Planning was performed on a mid-ventilation image acquired from a respiration-correlated CT scan. The blurring effect of tumour position variability was included in the dose calculation by convolution. CTV to PTV margins were calculated for variability in the cranio-caudal direction. A population margin of 9.0 +/- 0.7 mm was required to account for setup error and respiration in the study population without the use of image-guidance. The greatest mean margin reduction was introduced by the offline adaptive strategy. A daily online correction strategy produced a small reduction (1.6 mm) in the mean margin from the offline strategy. Adaptively correcting for an inter-fraction change in the respiratory pattern had little effect on margin size due to most patients having only small daily changes in the respiratory pattern. A daily online correction strategy would be useful for patients who exhibit large variations in the daily mean tumour position, while an offline adaptive strategy is more applicable to patients with less variation.

Fluoroscopy↗

Patient and staff dosimetry in vertebroplasty.

STUDY DESIGN: Eleven vertebroplasty operations were studied in terms of radiation dose. OBJECTIVE: Doses to patients and staff associated with vertebroplasty were measured. Occupational doses were compared with the annual dose limits, and the effectiveness of the used radiation protection means was estimated. Patient dose was estimated by means of both surface and effective dose, and the radiation-induced risk was evaluated. SUMMARY OF BACKGROUND DATA: Vertebroplasty is a recent minimally invasive technique for the restoration of vertebral body fractures. It involves fluoroscopic exposure, and so, it demands dose measurements for both patient and staff exposed to radiation. METHODS: Thermoluminescent dosimeters (TLDs) were placed on the medical personnel and the effective dose was derived. Slow films were placed to patients' skin to measure entrance surface dose. Furthermore, a Rando phantom loaded with TLDs was irradiated under conditions simulating vertebroplasty, in order to estimate effective dose to the patient. RESULTS: Mean fluoroscopy time was 27.7 minutes. Patient's mean skin dose was 688 mGy, while effective dose was calculated to be 34.45 mGy. It was estimated that the primary operator can perform about 150 vertebroplasty operations annually without exceeding the annual dose constraints, whereas occupational dose can be reduced by 76% using mobile shielding. CONCLUSIONS: Measures have to be taken to reduce patient's skin dose, which, in extreme cases, may be close to deterministic effects threshold. The highest dose rates, recorded during the procedure, were found for primary operator's hands and chest when no shielding was used.

Fluoroscopy↗

An inter-hospital comparison of patient dose based on clinical indications.

Patient dose is usually estimated for a single radiographic projection or computed tomography (CT) series. In this study, patient dose was calculated for predefined clinical indications (24 radiography, 11 CT). Members of the radiology staff of each of 11 hospitals were trained in dose measurement and calculation techniques. Based on clinical indications participants decided on imaging protocols and calculated cumulative effective dose for a complete examination. Effective dose ranged from <1 microSv to 0.6 mSv for examinations with radiographs and from 0.2 to 12 mSv for CT scans. Differences in the imaging protocols contributedd to a substantial variation in patient dose. For mammography, average glandular dose (AGD) was estimated for 32-, 53- and 90-mm compressed breast thicknesses, with a median value of 0.74, 1.74 and 3.40 mGy, respectively. The results presented here demonstrate that a pragmatic choice of dosimetry methods enables local staff to estimate effective dose. The inclusion of imaging protocols in the dose surveys provided a broader view on the variations in patient dose between hospitals.

Fluoroscopy↗